FDA Enforcement
Class II
Terminated
StoneBreaker Exhaust Line, Catalog number: SBA-EL; GPN: G52600; UDI: (01)00827002526006
Recall: Z-1592-2018
·
Reported May 9, 2018
Enforcement
- Recall Number
- Z-1592-2018
- Event ID
- 79731
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cook Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 9, 2018
- Initiation Date
- January 31, 2018
- Classification Date
- May 2, 2018
- Termination Date
- April 17, 2020
- Address
- 750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States
Description
StoneBreaker Exhaust Line, Catalog number: SBA-EL; GPN: G52600; UDI: (01)00827002526006
Reason
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Code Info
All Lots
Distribution
US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, UA, UY, XK, ZA
Quantity
760