FDA Enforcement
Class II
Ongoing
Identity Imprint PS Tibial Tray Size 4: Lot 540287
Recall: Z-1591-2024
·
Reported April 24, 2024
Enforcement
- Recall Number
- Z-1591-2024
- Event ID
- 94288
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Conformis, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 24, 2024
- Initiation Date
- March 11, 2024
- Classification Date
- April 17, 2024
- Address
- 600 Technology Park Dr # 400, Billerica, MA, 01821-4154, United States
Description
Identity Imprint PS Tibial Tray Size 4: Lot 540287
Reason
Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were swapped between the pouching and final pack.
Code Info
Catalog Number: TPS-301-T04S; UDI#: M572TPS301T04S011, Lot number 540287
Distribution
US Nationwide distribution in the states of FL, KS and NH.
Quantity
3 units