FDA Enforcement Class II Terminated

Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray system Product Usage: The Allura Xper ED series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, eg peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placement, embolization and thrombolysis. Cardiac imaging applications including diagnostics, interventional minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper ED series is compatible with a hybrid operating room.

Recall: Z-1589-2018 · Reported May 9, 2018

Enforcement

Recall Number
Z-1589-2018
Event ID
79770
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Electronics North America Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 9, 2018
Initiation Date
March 14, 2018
Classification Date
May 1, 2018
Termination Date
April 29, 2024
Address
3000 Minuteman Rd, N/A, Andover, MA, 01810-1032, United States

Description

Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray system Product Usage: The Allura Xper ED series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, eg peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placement, embolization and thrombolysis. Cardiac imaging applications including diagnostics, interventional minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper ED series is compatible with a hybrid operating room.

Reason

After continuous operation for more than one and a half days, the image on the large screen monitor may freeze for approximately 15 seconds after which the system will restore itself.

Code Info

Product name: Allura Xper R8.1 .25.0 Allura Xper R8.1 .25.1 Allura Xper R8.1 .25.5 Allura Xper R8.2.25.0 Allura Xper R8.2.25.5 Allura Xper R8.2.27 UNIQ Ri .0.10 UNIQ Ri .0.10.5 Product code: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 7322039, 722058, 722059.

Distribution

Worldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The product was distributed to the following countries: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Croatia, Cuba, Czech Republic, Denmark, Estonia, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Republic of Korea, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, Norway, Pakistan, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, and United Kingdom.

Quantity

682