AXIOM ARISTOS FX; Multipurpose Radiography System (MPRS) is a dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems. The MPRS allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities, and excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions. The intended use and indications for use of this modified device as described in its labeling have not changed from its predicate device.

Recall: Z-1583-2015 · Reported May 13, 2015

Enforcement

Recall Number
Z-1583-2015
Event ID
71029
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 13, 2015
Initiation Date
April 14, 2015
Classification Date
May 5, 2015
Termination Date
August 17, 2016
Address
51 Valley Stream Pkwy, N/A, Malvern, PA, 19355-1406, United States

Description

AXIOM ARISTOS FX; Multipurpose Radiography System (MPRS) is a dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems. The MPRS allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities, and excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions. The intended use and indications for use of this modified device as described in its labeling have not changed from its predicate device.

Reason

If the cable cart and the corrugated hose are unable to move freely, the cable holder could drop down and hit patients or system operators.

Code Info

Model Number of device: 7414803, Serial #s: 1008 1121 1016 1156 1157 1056 1043 1126 1076 1137 1088 1093 1030 1033 1029 1032 1100 1124 1139 1050 1058 1078 1127 1040 1035 1155 1143 1074 1109 1135 1007 1131 1110 1081 1064 1145 1146 1147 1059 1122 1140 1003 1083

Distribution

Nationwide Distribution.

Quantity

43