FDA Enforcement Class II Terminated

ViewRay System, Radiation Therapy System

Recall: Z-1580-2015 · Reported May 13, 2015

Enforcement

Recall Number
Z-1580-2015
Event ID
71050
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Viewray Incorporated
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
May 13, 2015
Initiation Date
January 15, 2015
Classification Date
May 5, 2015
Termination Date
September 23, 2015
Address
2 Thermo Fisher Way, N/A, Oakwood Village, OH, 44146-6536, United States

Description

ViewRay System, Radiation Therapy System

Reason

The software was not correctly using the RT (Radiation Therapy) to MR (Magnetic Resonance image) coordinate correction for non HFS (Head First Supine) patient orientations, resulting in slice mismatch error.

Code Info

M/N 10000, S/N 100 only

Distribution

US distribution to MO.

Quantity