FDA Enforcement Class II Ongoing

STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612

Recall: Z-1576-2023 · Reported May 17, 2023

Enforcement

Recall Number
Z-1576-2023
Event ID
92007
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Diagnostica Stago, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 17, 2023
Initiation Date
March 22, 2023
Classification Date
May 10, 2023
Address
5 Century Dr, N/A, Parsippany, NJ, 07054-4607, United States

Description

STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612

Reason

An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).

Code Info

UDI-DI: 036074505860; Serial Numbers: 7096527 7116678

Distribution

US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.

Quantity

2 analyers