FDA Enforcement
Class II
Ongoing
STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612
Recall: Z-1576-2023
·
Reported May 17, 2023
Enforcement
- Recall Number
- Z-1576-2023
- Event ID
- 92007
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Diagnostica Stago, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 17, 2023
- Initiation Date
- March 22, 2023
- Classification Date
- May 10, 2023
- Address
- 5 Century Dr, N/A, Parsippany, NJ, 07054-4607, United States
Description
STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612
Reason
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).
Code Info
UDI-DI: 036074505860; Serial Numbers: 7096527 7116678
Distribution
US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.
Quantity
2 analyers