FDA Enforcement Class II Terminated

1. Groshong 9.5 Dual-Lumen CV Catheter, Ref: 7726950, UDI: (01)00801741041570; 2. Groshong 9.5 F Dual-Lumen CV Catheter, REF: 7726954, UDI: (01)00801741041587

Recall: Z-1574-2020 · Reported April 1, 2020

Enforcement

Recall Number
Z-1574-2020
Event ID
84914
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Bard Peripheral Vascular Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 1, 2020
Initiation Date
October 1, 2019
Classification Date
March 26, 2020
Termination Date
February 18, 2022
Address
1625 W 3rd St Bldg 1, N/A, Tempe, AZ, 85281-2438, United States

Description

1. Groshong 9.5 Dual-Lumen CV Catheter, Ref: 7726950, UDI: (01)00801741041570; 2. Groshong 9.5 F Dual-Lumen CV Catheter, REF: 7726954, UDI: (01)00801741041587

Reason

Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.

Code Info

7726950 Lot # RECQ1106, RECR2195 7726954 Lot # RECU1837

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS,KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT,VA, VT, WA, WI, WV, WY and countries of Brazil, Canada, EMEA, Korea, Mexico, New Zealand.

Quantity

650 Catheters