ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844 ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System (Russia) 10658845 - ACUSON X700 Ultrasound System (Korea) Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesphageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Enforcement
- Recall Number
- Z-1572-2016
- Event ID
- 73886
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Visit
- Report Date
- May 4, 2016
- Initiation Date
- September 5, 2015
- Classification Date
- April 27, 2016
- Termination Date
- April 27, 2016
- Address
- 685 E Middlefield Rd, N/A, Mountain View, CA, 94043-4045, United States
Description
ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844 ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System (Russia) 10658845 - ACUSON X700 Ultrasound System (Korea) Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesphageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which requires a reboot to recover system operation.
Material (System) IVK Group Serial number: 10658844 ACUSON X700 356559 10658844 ACUSON X700 356733 10658844 ACUSON X700 356734 10658844 ACUSON X700 358714
Distributed in Germany, U.A.E., Hungary and Brazil.