FDA Enforcement Class II Terminated

ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844  ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System (Russia) 10658845 - ACUSON X700 Ultrasound System (Korea) Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesphageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Recall: Z-1572-2016 · Reported May 4, 2016

Enforcement

Recall Number
Z-1572-2016
Event ID
73886
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Visit
Report Date
May 4, 2016
Initiation Date
September 5, 2015
Classification Date
April 27, 2016
Termination Date
April 27, 2016
Address
685 E Middlefield Rd, N/A, Mountain View, CA, 94043-4045, United States

Description

ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844  ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System (Russia) 10658845 - ACUSON X700 Ultrasound System (Korea) Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesphageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Reason

Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which requires a reboot to recover system operation.

Code Info

Material (System) IVK Group Serial number: 10658844 ACUSON X700 356559 10658844 ACUSON X700 356733 10658844 ACUSON X700 356734 10658844 ACUSON X700 358714

Distribution

Distributed in Germany, U.A.E., Hungary and Brazil.

Quantity

4