FDA Enforcement Class II Terminated

Hardy Diagnostics GBS Detect Cat no. A300. Used for the isolation and detection by enhanced hemolysis of gamma-hemolytic (non-hemolytic) Group B Streptococcus.

Recall: Z-1571-2016 · Reported May 4, 2016

Enforcement

Recall Number
Z-1571-2016
Event ID
73688
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hardy Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
May 4, 2016
Initiation Date
March 21, 2016
Classification Date
April 26, 2016
Termination Date
October 25, 2016
Address
1430 W McCoy Ln, N/A, Santa Maria, CA, 93455-1005, United States

Description

Hardy Diagnostics GBS Detect Cat no. A300. Used for the isolation and detection by enhanced hemolysis of gamma-hemolytic (non-hemolytic) Group B Streptococcus.

Reason

Performance failure; poor hemolytic reaction of non or weakly hemolytic group B streptococcus (GBS) with the target QC organism, Streptococcus agalactiae (ATCC 13813) due to product deterioration.

Code Info

Lot H16054 and H15056.

Distribution

U.S. Distribution to the following states: AL, SC, FL, CT, VA, WI, MA, MI, WV, OH, IN, IL, NC, and TX.

Quantity

3,530 (353 pk/10)