FDA Enforcement Class I Ongoing

Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 12618, e) 20678, f) 20679, g) 20792, h) 60629, i) 12348, j) 11973

Recall: Z-1566-2023 · Reported May 24, 2023

Enforcement

Recall Number
Z-1566-2023
Event ID
91874
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
ICU Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 24, 2023
Initiation Date
March 22, 2023
Classification Date
May 16, 2023
Address
600 N Field Dr, N/A, Lake Forest, IL, 60045-4835, United States

Description

Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 12618, e) 20678, f) 20679, g) 20792, h) 60629, i) 12348, j) 11973

Reason

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

Code Info

All devices containing CSB BATTERY displaying the first two (2) characters as 22 or lower

Distribution

worldwide

Quantity

23815 units