FDA Enforcement Class II Ongoing

REF CH3397 Oncology Kit w/Spiros¿, Red Cap, Bag Spike w/Clave Additive Port LOT 5118610 - Product Usage: for use in the preparation and patient administration of cytotoxic medications.

Recall: Z-1565-2021 · Reported May 19, 2021

Enforcement

Recall Number
Z-1565-2021
Event ID
87703
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
ICU Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 19, 2021
Initiation Date
March 16, 2021
Classification Date
May 13, 2021
Address
951 Calle Amanecer, San Clemente, CA, 92673-6212, United States

Description

REF CH3397 Oncology Kit w/Spiros¿, Red Cap, Bag Spike w/Clave Additive Port LOT 5118610 - Product Usage: for use in the preparation and patient administration of cytotoxic medications.

Reason

Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.

Code Info

REF: CH3397 UDI: (01) 1 0840619 07367 8 (17) 251201 (30) 50 (10) 5118610 Lot Number: 5118610

Distribution

Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA and WA. The countries of Belgium, Canada, France, Italy, South Africa, Spain, United Kingdom.