VenaSeal Closure System
Enforcement
- Recall Number
- Z-1564-2021
- Event ID
- 87619
- Classification
- Class III
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medtronic Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 19, 2021
- Initiation Date
- March 23, 2021
- Classification Date
- May 12, 2021
- Address
- 4600 Nathan Ln N, Plymouth, MN, 55442-2890, United States
Description
VenaSeal Closure System
Firm received increased number of reports that VenaSeal dispensing systems are not effectively advancing the adhesive forward through the delivery catheter.
Model Number SP-101: GTIN 20643169986279, Serial Numbers: 54143, 54145, 54146, 54147, 54148, 54149, 54150, 54151, 54152, 54396, 54397, 54398, 54399, 54417, 54418, 54419, 54420, 54618, 54619, 54620, 54621, 54622, 54771, 54772, 54773, 54774, 54775, 54776, 54794, 54795, 54796, 54924, 54925, 54926, 54927, 54928, 54929, 55027, 55028, 55029, 55030, 55031, 55032, 55068, 55069, 55070, 55071, 55072, 55073, 55113, 55114, 55115, 55116, 55437, 55438, 55439, 55440, 55441, 55442, 55483, 55484, 55485, 55486, 55487, 55488, 55594, 55595, 55596, 55597, 55598, 55599, 55675, 55676, 55677, 55678, 55683, 55684, 55775, 55776, 55777, 55778, 55790, 55791, 55792, 55793, 55794, 55852, 55853, 55854, 55855, 55856, 55948, 55949, 55951, 55952, 55953, 55991, 55992, 55993, 55994, 55995, 56109, 56110, 56111, 56123, 56159, 56160, 56161, 56162, 56163, 56189, 56193, 56195, 56196, 56216, 56218, 56219, 56387, 56388, 56389, 56391, 56464, 56465, 56466, 56467, 56468, 56528, 56529, 56530, 56531, 56532, 56533, 56534, 56661, 56662, 56663, 56664, 56741, 56742, 56829, 56830, 56831, 56832, 56833, 56846, 56847, 56848, 56928, 56929, 56930, 56931, 56932, 56933, 56935, 57063, 57177, 57178, 57179, 57226, 57227, 57268, 57269, 57270, 57271, 57272, 57273, 57439, 57440, 57560, 57561, 57562, 57563, 57564, 57565, 57567, 57884, 57885, 57947, 57948, 58257, 58258, 58363, 58364, 58564, 58566, 58567, 58568, 58897, 58898, 58899, 59249, 59337, 59338, 59339, 59909, 59910, 59911, 60264, 60265, 60682, 60683, 60684, 60685, 60763, 60764, 60765, 60774, 60775, 60776, 60876, 60877, 60880, 60881, 60882, 60883, 60884, 60886, 60887, 60888, 60889, 60890, 60891, 60892, 60893, 60894, 60895, 60896, 60897, 60898, 60899, 60900, 61189, 61190, 61191, 61192, 61193, 61254, 61257, 61262, 61264, 61312, 61313, 61314, 61315, 61598, 61599, 61600, 61605, 61606, 61607, 61612, 61613, 61617, 61925, 61928, 61935, 61936, 61937, 61944, 61946, 61953, 61960, 53106, 53107, 53388, 53449, 54420, 54618, 54929, 55028 Model Number VS-402: GTIN 00613994384089, Serial Numbers: 53830, 54210, 54211, 54421, 55491 GTIN 00643169437937, Serial Numbers: 53841 GTIN 00643169931459, Serial Numbers: 53928 GTIN 20643169986262, Serial Numbers: 53821, 53834, 53835, 53842, 53843, 53845, 53894, 53896, 53900, 53926, 53927, 53929, 53938, 54141, 54212, 54214, 54215, 54216, 54217, 54218, 54220, 54221, 54222, 54287, 54288, 54289, 54291, 54292, 54318, 54320, 54321, 54323, 54325, 54360, 54362, 54363, 54367, 54371, 54372, 54393, 54395, 54422, 54549, 54550, 54552, 54555, 54558, 54559, 54567, 54568, 54695, 54696, 54697, 54698, 54777, 54779, 54780, 54791, 54792, 54793, 54830, 54831, 54834, 55076, 55077, 55110, 55111, 55112, 55227, 55357, 55443, 55444, 55489, 55601, 55602, 55679, 55680, 55686, 55687, 55779, 55782, 55796, 55797, 55860, 55861, 55955, 55959, 55962, 55963, 55971, 56113, 56115, 56116, 56164, 56165, 56166, 56167, 56197, 56198, 56199, 56200, 56201, 56220, 56221, 56222, 56384, 56385, 56386, 56469, 56470, 56471, 56472, 56473, 56535, 56536, 56537, 56641, 56736, 56737, 56739, 56740, 56818, 56819, 56820, 56821, 56822, 56823, 56824, 56825, 56826, 56827, 56840, 56841, 56842, 56843, 56844, 56845, 56937, 56938, 56939, 56940, 56941, 56942, 56943, 56944, 56945, 56946, 56947, 56948, 56949, 56950, 56951, 56952, 56953, 56954, 56955, 57077, 57082, 57180, 57181, 57182, 57183, 57184, 57185, 57186, 57228, 57229, 57230, 57231, 57232, 57233, 57234, 57235, 57236, 57237, 57274, 57275, 57276, 57277, 57278, 57279, 57280, 57281, 57282, 57283, 57284, 57285, 57286, 57287, 57288, 57289, 57290, 57291, 57296, 57297, 57298, 57299, 57568, 57569, 57570, 57571, 57572, 57573, 57574, 57575, 57576, 57577, 57578, 57579, 57580, 57581, 57582, 57583, 57584, 57586, 57587, 57588, 57589, 57590, 57591, 57592, 57593, 57594, 57595, 57596, 57597, 57598, 57599, 57600, 57601, 57602, 57603, 57604, 57605, 57606, 57607, 57608, 57886, 57887, 57888, 57889, 57949, 57950, 57951, 57952, 57953, 57954, 57955, 57962, 57963, 57964, 57965, 57966, 57967, 57968, 57969, 57970, 57971, 57972, 58057, 58058, 58059, 58060, 58061, 58062, 58081, 58082, 58083, 58084, 58085, 58101, 58102, 58103, 58104, 58105, 58107, 58108, 58109, 58110, 58111, 58112, 58113, 58193, 58194, 58195, 58259, 58260, 58261, 58262, 58356, 58357, 58358, 58359, 58360, 58378, 58379, 58380, 58381, 58382, 58383, 58384, 58385, 58386, 58387, 58388, 58389, 58415, 58416, 58417, 58418, 58419, 58420, 58421, 58422, 58423, 58424, 58425, 58426, 58427, 58428, 58429, 58430, 58431, 58432, 58433, 58434, 58435, 58436, 58437, 58438, 58439, 58440, 58441, 58442, 58569, 58570, 58571, 58572, 58573, 58602, 58603, 58604, 58605, 58609, 58610, 58611, 58612, 58613, 58614, 58615, 58616, 58617, 58618, 58619, 58620, 58621, 58622, 58623, 58624, 58625, 58626, 58627, 58628, 58629, 58630, 58631, 58632, 58633, 58634, 58635, 58636, 58637, 58638, 58639, 58640, 58641, 58642, 58643, 58644, 58645, 58646, 58647, 58839, 58901, 58902, 58903, 58904, 58905, 58906, 58907, 58908, 58988, 58989, 58990, 58991, 58992, 58993, 58994, 58995, 58996, 58997, 58998, 58999, 59000, 59001, 59002, 59003, 59004, 59005, 59006, 59007, 59008, 59009, 59010, 59011, 59012, 59013, 59014, 59015, 59016, 59017, 59018, 59019, 59020, 59021, 59022, 59023, 59151, 59234, 59251, 59252, 59253, 59254, 59255, 59256, 59257, 59258, 59259, 59260, 59261, 59262, 59263, 59264, 59265, 59266, 59267, 59268, 59269, 59270, 59271, 59272, 59273, 59274, 59275, 59276, 59277, 59278, 59279, 59280, 59281, 59282, 59283, 59284, 59285, 59286, 59343, 59344, 59345, 59349, 59356, 59357, 59358, 59359, 59360, 59361, 59362, 59363, 59364, 59365, 59366, 59367, 59368, 59584, 59585, 59586, 59690, 59912, 59913, 59914, 59915, 59916, 59954, 59955, 59956, 59957, 59958, 59959, 59960, 59961, 59962, 59963, 59964, 59965, 59966, 59967, 59968, 59969, 59970, 59971, 59972, 59973, 59974, 60124, 60266, 60267, 60268, 60269, 60270, 60271, 60272, 60273, 60274, 60275, 60276, 60277, 60278, 60279, 60280, 60281, 60282, 60283, 60687, 60688, 60689, 60690, 60691, 60692, 60693, 60694, 60695, 60766, 60767, 60768, 60777, 60778, 60779, 60780, 60781, 60782, 60783, 60784, 60785, 60786, 60787, 60788, 60793, 60794, 60795, 60796, 60797, 60798, 60799, 60800, 60801, 60812, 60813, 60814, 60815, 60816, 60818, 60819, 60820, 60821, 60822, 60823, 60824, 60825, 60826, 60828, 60829, 60830, 60831, 60832, 60833, 60834, 60878, 60879, 60901, 60902, 60903, 60904, 60905, 60906, 60907, 60908, 60909, 61187, 61188, 61246, 61247, 61248, 61249, 61250, 61251, 61252, 61253, 61258, 61259, 61260, 61261, 61265, 61266, 61267, 61268, 61269, 61270, 61271, 61272, 61273, 61274, 61275, 49967, 50226, 50229, 50431, 50547, 50651, 50816, 50832, 50833, 50972, 51017, 51076, 51077, 51178, 51195, 51307, 51310, 51461, 51509, 51590, 51657, 51659, 51660, 51752, 51755, 51757, 51768, 51794, 51795, 51856, 51859, 51872, 51908, 51909, 51910, 51958, 51961, 51982, 52028, 52029, 52128, 52129, 52130, 52138, 52139, 52217, 52222, 52224, 52225, 52226, 52274, 52275, 52293, 52294, 52297, 52372, 52373, 52412, 52413, 52414, 52415, 52416, 52417, 52420, 52426, 52427, 52429, 52438, 52441, 52445, 52447, 52448, 52449, 52454, 52483, 52485, 52487, 52488, 52489, 52490, 52491, 52504, 52505, 52506, 52599, 52600, 52601, 52602, 52639, 52640, 52641, 52642, 52646, 52647, 52648, 52649, 52706, 52707, 52708, 52718, 52719, 52720, 52721, 52722, 52723, 52728, 52729, 52737, 52738, 52739, 52740, 52741, 52790, 52791, 52792, 52793, 52794, 52795, 52796, 52797, 52801, 52802, 52803, 52804, 52805, 52806, 52815, 52816, 52817, 52818, 52819, 52820, 52821, 52823, 52829, 52832, 52833, 52834, 52835, 52836, 52837, 52838, 52839, 52895, 52896, 52897, 52898, 52899, 52900, 52901, 52902, 52903, 52904, 52905, 52910, 52911, 52912, 52913, 52914, 52915, 52916, 52917, 52918, 52938, 52939, 52940, 52941, 52942, 52943, 52944, 52945, 53025, 53026, 53042, 53043, 53044, 53051, 53052, 53053, 53054, 53055, 53056, 53057, 53058, 53060, 53061, 53062, 53063, 53064, 53098, 53099, 53100, 53101, 53170, 53171, 53172, 53173, 53174, 53175, 53176, 53186, 53187, 53188, 53189, 53190, 53197, 53198, 53199, 53200, 53201, 53202, 53212, 53213, 53214, 53215, 53216, 53217, 53254, 53255, 53256, 53257, 53258, 53259, 53260, 53261, 53262, 53263, 53264, 53265, 53266, 53267, 53268, 53317, 53318, 53319, 53320, 53321, 53322, 53323, 53324, 53325, 53326, 53327, 53328, 53339, 53459, 53460, 53461, 53463, 53464, 53465, 53466, 53467, 53468, 53539, 53540, 53541, 53542, 53543, 53562, 53563, 53565, 53566, 53567, 53584, 53585, 53586, 53587, 53588, 53637, 53638, 53639, 53640, 53652, 53653, 53654, 53655, 53656, 53703, 53712, 53713, 53714, 53715, 53716, 53718, 53719, 53720, 53721, 53722, 53814, 53815, 53816, 53817, 53818, 53819, 53820, 53821, 53830, 53831, 53832, 53833, 53834, 53835, 53841, 53842, 53843, 53844, 53845, 53846, 53890, 53895, 53896, 53897, 53898, 53899, 53900, 53901, 53925, 53926, 53927, 53928, 53929, 53931, 53936, 53937, 53938, 53939, 53940, 53941, 54052, 54053, 54054, 54140, 54141, 54209, 54210, 54211, 54212, 54214, 54215, 54216, 54217, 54218, 54219, 54220, 54221, 54287, 54288, 54289, 54290, 54291, 54318, 54320, 54321, 54322, 54323, 54324, 54325, 54326, 54327, 54359, 54360, 54361, 54362, 54363, 54364, 54365, 54366, 54367, 54368, 54369, 54370, 54371, 54372, 54373, 54392, 54393, 54394, 54395, 54421, 54422, 54423, 54424, 54451, 54452, 54453, 54454, 54455, 54456, 54457, 54549, 54550, 54551, 54552, 54553, 54554, 54555, 54556, 54557, 54558, 54559, 54567, 54568, 54695, 54696, 54697, 54698, 54777, 54778, 54779, 54780, 54781, 54782, 54783, 54790, 54791, 54792, 54793, 54830, 54831, 54832, 54833, 54834, 55074, 55075, 55076, 55077, 55107, 55108, 55109, 55110, 55111, 55112, 55224, 55225, 55226, 55227, 55228, 55229, 55230, 55231, 55353, 55354, 55355, 55356, 55357, 55358, 55359, 55360, 55361, 55362, 55364, 55365, 55443, 55444, 55445, 55489, 55490, 55491, 55600, 55601, 55602, 55679, 55680, 55681, 55682, 55685, 55686, 55687, 55779, 55780, 55781, 55782, 55795, 55796, 55797, 55858, 55859, 55860, 55861, 55862, 55955, 55959, 55962, 55963, 55970, 56114, 56115, 56164, 56165, 56166, 56167, 56197, 56198, 56199, 56200, 56202, 56384, 56469, 56470, 56473 Model Number VS-404: GTIN 00643169437975, Serial Numbers: 61677 GTIN 00673978070469, Serial Numbers: 62015 GTIN 00673978070872, Serial Numbers: 62006 GTIN 00763000240776, Serial Numbers: 61475, 61673, 61674, 61678, 61696, 61933, 61934, 61938, 61939, 61940, 61941, 61943, 61947, 62013, 62022 GTIN 00763000279769, Serial Numbers: 61675 Please note that VS-403 is not approved in the U.S. and has not been distributed outside of the United States; the scope of this corrective action will include this product as it is launched in OUS regions beginning Summer 2021.
Distributed nationwide to MA, NY, VT, NH, ME
154,152 units