FDA Enforcement Class II Terminated

Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Magellan Catheters and 3rd-party guide wires performing smoothly during system operation.

Recall: Z-1564-2016 · Reported May 4, 2016

Enforcement

Recall Number
Z-1564-2016
Event ID
73677
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hansen Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 4, 2016
Initiation Date
March 25, 2016
Classification Date
April 25, 2016
Termination Date
May 6, 2016
Address
800 E Middlefield Rd, N/A, Mountain View, CA, 94043-4030, United States

Description

Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Magellan Catheters and 3rd-party guide wires performing smoothly during system operation.

Reason

Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

Code Info

Lot number 20151117

Distribution

Nationwide Distribution to OH, TX. FL, CA and GA.

Quantity

8 kits