Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,L MDTGXC4JXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XL MDTGXC4JXXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XX MDTGXP4JL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,L MDTGXP4JXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTGXP4JXXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XX MDTL1002862EF DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTZ1000893AJ DBD-GOWN,SRG,PNL CVR, GREEN,SNPS,XL
Enforcement
- Recall Number
- Z-1563-2026
- Event ID
- 98430
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 18, 2026
- Initiation Date
- January 28, 2026
- Classification Date
- March 12, 2026
- Address
- 3 Lakes Dr, N/A, Northfield, IL, 60093-2753, United States
Description
Medline Medcrest Surgical Gowns: MDTGXC4J5XL DBQ-GOWN,SURG,XALT,L4, CRTCL CVR, GRN, 5 MDTGXC4JL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,L MDTGXC4JXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XL MDTGXC4JXXL DBD-GOWN,SURG,XALT,L4,CRTCL CVR,GREEN,XX MDTGXP4JL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,L MDTGXP4JXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTGXP4JXXL DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XX MDTL1002862EF DBD-GOWN,SURG,XALT,L4,PANEL CVR,GREEN,XL MDTZ1000893AJ DBD-GOWN,SRG,PNL CVR, GREEN,SNPS,XL
Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
UDI-DI (each/case) 10197344014002 20197344014009 10888277189188 40888277189189 10884389189191 40884389189192 10884389193297 40884389193298 10888277189218 40888277189219 10884389189221 40884389189222 10193489040111 20193489040118 10884389189221 40884389189222 10193489021820 20193489021827 All lots affected
Domestic distribution nationwide US. International distribution to Panama, United Arab Emirates, Virgin Islands, Saudi Arabia, India, Chile, and Canada.
104,004