DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb Holder, Ankle, Fixed, Quick Release, REF M8125; (3) DeRoyal Security Cuffs, Wrist and Ankle, Universal, REF M8136; and (4) DeRoyal Security Cuffs, Heavy Duty Vertical Psychiatric, University, REF M8137.
Enforcement
- Recall Number
- Z-1562-2022
- Event ID
- 90629
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- DeRoyal Industries Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 31, 2022
- Initiation Date
- July 15, 2022
- Classification Date
- August 22, 2022
- Termination Date
- August 6, 2024
- Address
- 200 Debusk Ln, Powell, TN, 37849-4703, United States
Description
DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb Holder, Ankle, Fixed, Quick Release, REF M8125; (3) DeRoyal Security Cuffs, Wrist and Ankle, Universal, REF M8136; and (4) DeRoyal Security Cuffs, Heavy Duty Vertical Psychiatric, University, REF M8137.
The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure.
(1) M8124 - Lot 57372551 / exp 06/06/2032; Lot 57285721 / exp 05/22/2032; Lot 57369651 / exp 06/06/2032; Lot 57497652 / exp 6/27/2032; GTIN: 00749756780728; (2) M8125 - Lot 5746744 / exp 6/20/2032; Lot 57419195 / exp 6/13/2032; Lot 57420031 / exp 6/14/2032; GTIN: 00749756780735; (3) M8136 - Lot 57501121 / exp 6/27/2032; GTIN: 00749756837828; (4) M8137 - Lot 57462151 / exp 6/20/2032; GTIN: 00749756849234.
US Nationwide distribution in the states of FL, IN, NV, RI, TN, TX, and WV.
36 eaches and 7 pairs