FDA Enforcement Class II Terminated

Radionuclide Radiation Therapy System Product The product is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.

Recall: Z-1561-2016 · Reported May 4, 2016

Enforcement

Recall Number
Z-1561-2016
Event ID
73761
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Elekta, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 4, 2016
Initiation Date
April 19, 2016
Classification Date
April 22, 2016
Termination Date
July 16, 2021
Address
400 Perimeter Center Ter NE Ste 50, N/A, Atlanta, GA, 30346-1227, United States

Description

Radionuclide Radiation Therapy System Product The product is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.

Reason

The latches of the frame adapter can be locked even if the locating pins of the frame adapter is not inserted into the corresponding holes in the coordinate frame.

Code Info

Perfexion - 14753-099/6052, 12579-007/6250, 10394-LGK01/6144, 11920-C5210/6088, 17397-001/6158, 105528-001/6122, 14857-001/6028, 12236-006/6097, 12473-003/6005, 14857-001/6028, 16867-001/6166, 12668-003/6145 10208-003/6215, 10698-001/6237, 10979-002/6213, 10316-500/6210, 21061-001/6195, 20364-001/6103, 11720-002/6143, 12620-002/6120, 10611-001/6081, 12846-004/6064, 11524-002/8001, 11420-001/6212, 11420-001/6228, 14983-001/6076, 11083-HOM001/6070, 10247-002/6136,10684-002/6090,11868-004/6002, 10795-K08002/6063, 20354-ENS/6160, 11792-005/6077, 11291-001/6018, 10326-003/6105, 18630-001/6031, 17322-001/6121, 30000656-002/6227, 12629-003/6026, 10924-001/6089, 11154-003/6050, 20438-003/6013 17522-001/6214, 17525-002/6038, 11414-LGK002/6205, 11940-LGK002/6181, 10514-LGK002/6183, 11639-LGK001/6042, 11390-LGK001/6109, 10416-LGK001/6125, 11670-LGK001/6133, 11990-LGK001/6118, 10648-LGK001/6098, 15104-LGK001/6080, 10654-LGK001/6096, 10959-LGK001/6132, 10997-LGK001/6087, 15748-LGK001/6106, 11149-LGK002/6206, 11986-LGK002/6218, 10929-LGK002/6175, 10879-LGK002/6234, 11434-LGK002/6255, 11991-LGK002/6198, 13480-LGK002/6148, 10425-LGK002/6254, 10583-LGK002/6161, 11563-LGK002/6162, 10103-LGK02/6149, 10541-LGK002/6142, 10848-LGK001/6108, 11433-LGK001/6116, 14900-LGK001/6058, 15045-LGK001/6091 11784-LGK001/6055, 30000732-001/6223, 40017-002/6170, 12085-006/6163, 12109-002/6169, 12959-002/6033, 10209-001/6046, 10960-005/6193, 0002718-001/6113, 30001751-WOJ701/6068, 12451-001/6019, 14878-003/6067, 105618-001/6110, 19634-001/6155, 16647-001/6187, 14891-001/6104, 30001527-001/6231, 12028-LGK002/6217, 11586-001/6157, 10728-001/6062, 11999-001/6126, 15416-001/6086, 10459-001/6123, 10158-001/6177, 10234-002/6111, 40023-001/6178, 10444-001/6085, 17033-003/6043, 11530-001/6101, 11553-001/6024, 10779-07/6069, 13465-CHUV03/6094, 10424-001/6182, 13370-002/6075, 12214-003/6007, 30000655-001/6199, 10152-001/6159, 11550-001/6065, 10118-001/6230, 12766-001/6112, 13160-001/6039, 11042-001/6003, 12270-001/6017, 18062-001/6146, 11710-004/6141, 11820-006/8006, 11570-004/6220, 11127-009/6219, 12164-004/6221, 10434-004/6244, 10542-012/6238, 13826-002/6239, 10567-041/6191, 98300-003/6140, 10567-037/6119, 14569-001/6135, 10643-005/6117, 11798-009/6060, 17357-001/6130, 11813-003/6034, 10356-007/6079, 10567-045/6232, 10567-032/6074, 10567-030/6036, 10567-025/6029, 11293-004/6004, 12269-002/6082, 11772-004/6048, 10164-004/6009, 10349-003/6066, 10567-026/6011, 10560-003/6127, 11073-008/6020, 10567-036/6102, 14236-001/6114, 10567-035/6092, 10379-002/6061, 11030-006/6030, 13050-001/6037, 10909-009/6010, 11714-003/6072, 12985-003/6025, 14030-002/6015, 10567-031/6054, 10064-008/6021, 13558-001/6099, 10567-038/6107, 11939-003/6008, 10591-002/6083, 10567-024/6014, 11848-003/6095, 11624-003/6078, 10383-007/6100, 11395-010/6032, 11198-005/6124, 10982-003/6053, 11256-001/6012, 10905-004/6016, 10387-011/6093, 11138-004/6115, 10524-003/6073, 10908-003/6045, 10762-005/6057, 10193-007/6056, 10673-016/6006, 11007-005/6035, 11791-006/6022, 11377-003/6186, 12752-004/6242, 11039-006/6188, 10597-002/6167, 17539-001/6185, 14153-001/6197, 30000118-001/6240, 10567-039/6180, 15053-001/6179, 11731-005/6200, 11090-005/6165, 11077-007/6207, 30000171-002/6172, 10760-006/6173, 11733-001/6192, 10817-009/6204, 13514-001/6211, 10877-GK01/6151, 10436-009/6150, 11396-003/6176, 10917-004/6128, 11373-002/6184, 10567-042/6129, 10567-043/6203, 12614-003/6139, 10297-006/6171, 11347-003/6134 LEKSELL GAMMA KNIFE ICON - 11868-008/6131, 11290-ENS1/6174, 12027-EZT-G1/6041, 10444-LGK004/8003, 11352-001/6044, 10622-005/6194, 10886-013/6027, 11201-004/6047, 11834-003/6222

Distribution

Worldwide Distribution - US Nationwide in the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WV, WI, and countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom

Quantity

217