FDA Enforcement Class II Terminated

Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.

Recall: Z-1560-2020 · Reported April 1, 2020

Enforcement

Recall Number
Z-1560-2020
Event ID
84891
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Dexcom Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Other
Report Date
April 1, 2020
Initiation Date
February 3, 2020
Classification Date
March 25, 2020
Termination Date
January 20, 2022
Address
6340 Sequence Dr, N/A, San Diego, CA, 92121-4356, United States

Description

Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.

Reason

The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.

Code Info

Device Listing Number: D189069 Model: STS-GL-041 Description: G4/G5 Sensor Kit 4 Pack GUDID Registered GTIN: 30386270000010 Lot Number: All Model: STS-RX-041 Description: G4/G5 Sensor Kit 4 Pack Retail GUDID Registered GTIN: 20386270000013 Lot Number: All Model: STS-MC-004 Description: G4/G5 Sensor Kit 4 Pack Medicare GUDID Registered GTIN: 00386270000637 Lot Number: All

Distribution

World-wide Distribution

Quantity

3091 Systems