Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities.
Enforcement
- Recall Number
- Z-1559-2013
- Event ID
- 65438
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hitachi Medical Systems America Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 26, 2013
- Initiation Date
- January 31, 2013
- Classification Date
- June 19, 2013
- Termination Date
- September 20, 2013
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087-2371, United States
Description
Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities.
Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.
The following serial numbers are associated with this recall: Echelon - V061, V062, and V065 Oasis - M009, M017, M067, M071, M084, M106-M108, M110-M141 inclusive, and M952 Echelon Oval - Y001-Y005 inclusive
Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, IL, IN, KS, KY, MD, NC, ND, NY, OH, OR, PA, TX, VA, WA, and WI. One (1) system unit was also distributed to Mexico.
45 system units