Philips Healthcare DigitalDiagnost System X-Ray
Enforcement
- Recall Number
- Z-1554-2015
- Event ID
- 70916
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 27, 2015
- Initiation Date
- February 9, 2015
- Classification Date
- May 21, 2015
- Termination Date
- June 14, 2016
- Address
- 3000 Minuteman Rd, N/A, Andover, MA, 01810-1032, United States
Description
Philips Healthcare DigitalDiagnost System X-Ray
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Digital Diagnost Stationary X-ray System Release 4.0.0, 4.0.1, 4.0.2
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.
54 DigitalDiagnost