FDA Enforcement Class II Ongoing

BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 155140-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 155143-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 155144-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 155145-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 155157-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P / 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 155153-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 155155-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 155147-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 155149-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 155150-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 169-114 BD PYXIS MEDBANK MINI CR-4HH-P / 169-115 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 169-83 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 169-85 BD PYXIS MEDBANK TWR MN CR-14HH-P / 169-86 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 169-89 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 169-91 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /169-93 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 169-95 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 169-99 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 138913-01 BD PYXIS MEDBANK MINI CR-4HH-P / 138911-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P / 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P / 139002-01 BD PYXIS MEDBANK MN 200 CR-4HH-P / 139000-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P / 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P / 139018-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P / 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P / 139021-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P / 139022-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 138975-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 138976-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 138977-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 139085-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 138964-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 138965-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 138970-01 BD Pyxis QFill Replenishment Station / 138904-01 KIT BD PYXIS QFILL REPLENISHMENT STATION / 155197-01 MEDBANK UPOS211 AIO WIN10 QFILL 1.5 / 155215-05 OPT,MEDBANK,HD,WIN10,QFILL,CR / 131944-04 The BD Pyxis" MedBank" CUBIE" Replenishment Station is an Automated Dispensing Cabinet (ADC) that is intended to be used by a qualified and authorized user to replenish medications and supplies to a BD Pyxis" MedBank" cabinet.

Recall: Z-1550-2025 · Reported April 16, 2025

Enforcement

Recall Number
Z-1550-2025
Event ID
96310
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
CareFusion 303, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 16, 2025
Initiation Date
March 11, 2025
Classification Date
April 9, 2025
Address
10020 Pacific Mesa Blvd, N/A, San Diego, CA, 92121-4386, United States

Description

BD Pyxis MedBank CUBIE Replenishment Station Product / Material ID BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 155140-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 155143-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 155144-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 155145-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 155157-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P / 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 155153-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 155155-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 155147-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 155149-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 155150-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 169-114 BD PYXIS MEDBANK MINI CR-4HH-P / 169-115 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 169-83 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 169-85 BD PYXIS MEDBANK TWR MN CR-14HH-P / 169-86 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 169-89 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 169-91 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /169-93 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 169-95 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 169-99 BD PYXIS MEDBANK MINI CR-2HH-1FH-P / 138913-01 BD PYXIS MEDBANK MINI CR-4HH-P / 138911-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P / 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P / 139002-01 BD PYXIS MEDBANK MN 200 CR-4HH-P / 139000-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P / 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P / 139018-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P / 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P / 139021-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P / 139022-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P / 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P / 138975-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH / 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P / 138976-01 BD PYXIS MEDBANK TWR MN CR-14HH-P / 138977-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM / 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH / 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM / 139085-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P / 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P / 138964-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P / 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P /138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P / 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P / 138965-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P / 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P / 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P / 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P / 138970-01 BD Pyxis QFill Replenishment Station / 138904-01 KIT BD PYXIS QFILL REPLENISHMENT STATION / 155197-01 MEDBANK UPOS211 AIO WIN10 QFILL 1.5 / 155215-05 OPT,MEDBANK,HD,WIN10,QFILL,CR / 131944-04 The BD Pyxis" MedBank" CUBIE" Replenishment Station is an Automated Dispensing Cabinet (ADC) that is intended to be used by a qualified and authorized user to replenish medications and supplies to a BD Pyxis" MedBank" cabinet.

Reason

Due to a software issue, when attempting to restock the automated dispensing cabinet, user have received the banner message "Stocking window has passed, or item has been removed from purchase order. Please contact Pharmacy for assistance". When this occurs, the medication cannot be restocked and can potentially lead to delays in accessing medication for patients.

Code Info

Product / Material ID / GTIN OPT,MEDBANK,HD,WIN10,QFILL,CR 131944-04 N/A BD Pyxis QFill Replenishment Station 138904-01 N/A BD PYXIS MEDBANK MINI CR-4HH-P 138911-01 N/A BD PYXIS MEDBANK MINI CR-2HH-1FH-P 138913-01 N/A BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 138951-01 N/A BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 138964-01 N/A BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 138965-01 N/A BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 138966-01 N/A BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 138967-01 N/A BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 138968-01 N/A BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 138969-01 N/A BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 138970-01 N/A BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 138971-01 N/A BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 138972-01 N/A BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 138973-01 N/A BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 138974-01 N/A BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 138975-01 N/A BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 138976-01 N/A BD PYXIS MEDBANK TWR MN CR-14HH-P 138977-01 N/A BD PYXIS MEDBANK MN 200 CR-4HH-P 139000-01 N/A BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P 139001-01 N/A BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P 139002-01 N/A BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P 139018-01 N/A BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P 139019-01 N/A BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P 139020-01 N/A BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P 139021-01 N/A BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P 139022-01 N/A BD PYXIS MEDBANK TWR MN CR-2HH-6FH 139072-01 N/A BD PYXIS MEDBANK TWR MN CR-12HH-1FH 139073-01 N/A BD PYXIS MEDBANK TWR MN CR-2HH-6FM 139085-01 N/A BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 155140-01 10885403504440 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 155141-01 10885403503580 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 155142-01 10885403504457 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 155143-01 10885403504464 BD PYXIS MEDBANK TWR MN CR-14HH-P 155144-01 10885403504471 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 155145-01 10885403504488 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 155146-01 10885403504754 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 155147-01 10885403504761 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 155148-01 10885403504778 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 155149-01 10885403504785 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 155150-01 10885403504792 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 155151-01 10885403504709 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 155152-01 10885403504716 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 155153-01 10885403504723 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 155154-01 10885403504730 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 155155-01 10885403504747 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 155156-01 10885403504624 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 155157-01 10885403504631 KIT BD PYXIS QFILL REPLENISHMENT STATION 155197-01 N/A MEDBANK UPOS211 AIO WIN10 QFILL 1.5 155215-05 N/A BD PYXIS MEDBANK MINI CR-2HH-1FH-P 169-114 10885403512544 BD PYXIS MEDBANK MINI CR-4HH-P 169-115 10885403512544 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 169-82 10885403512568 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 169-83 10885403512568 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 169-84 10885403512568 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 169-85 10885403512568 BD PYXIS MEDBANK TWR MN CR-14HH-P 169-86 10885403512568 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 169-87 10885403512568 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 169-88 10885403512568 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 169-89 10885403512568 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 169-90 10885403512568 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 169-91 10885403512568 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 169-92 10885403512568 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 169-93 10885403512568 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 169-94 10885403512568 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 169-95 10885403512568 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 169-96 10885403512568 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 169-97 10885403512568 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 169-98 10885403512568 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 169-99 10885403512568

Distribution

U.S.: AL,AZ, CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, PA, TN, TX, UT, VA, WA, WI and WV. O.U.S.: None

Quantity

404 systems