2008 Series Hemodialysis Machines: 2008T with the following product code and description: 190395 2008K at Home System - 1st Gen. -¿ Canada; 190828 2008K at Home System - 2nd Gen - U.S.; 190904 2008K at Home System w/ Bibag; 190828 2008KatHOME HEMODIALYSIS SYSTEM; 190395 2008KatHOME MACHINE, SHORT CAB, OLC/DP, HP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Enforcement
- Recall Number
- Z-1550-2015
- Event ID
- 70574
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Fresenius Medical Care Holdings, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 6, 2015
- Initiation Date
- February 20, 2015
- Classification Date
- April 29, 2015
- Termination Date
- May 18, 2021
- Address
- 920 Winter St, Waltham, MA, 02451-1521, United States
Description
2008 Series Hemodialysis Machines: 2008T with the following product code and description: 190395 2008K at Home System - 1st Gen. -¿ Canada; 190828 2008K at Home System - 2nd Gen - U.S.; 190904 2008K at Home System w/ Bibag; 190828 2008KatHOME HEMODIALYSIS SYSTEM; 190395 2008KatHOME MACHINE, SHORT CAB, OLC/DP, HP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Serial Numbers: 0T0S100031- 101358, 1T0S101359- 127938, 2T0S109489- 118130, 3T0S118131- 126163, 4T0S126164- 135408, 9T0S100002, 9T0S100004, 9T0S100007- 100010, 9T0S100012, 9T0S100014- 100016
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)