Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.
Enforcement
- Recall Number
- Z-1547-2013
- Event ID
- 65210
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 26, 2013
- Initiation Date
- April 29, 2013
- Classification Date
- June 14, 2013
- Termination Date
- May 6, 2015
- Address
- 56 E Bell Dr, Warsaw, IN, 46582-6989, United States
Description
Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.
Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su
Catalog number: 14-442021 and lot number:757910
Worldwide Distribution - USA Nationwide: CA, TX, NY, PA, NJ, FL, OH, MI, CO, SD, IN, MT, GA, and WI. and the countries of Netherlands, Costa Rica, Japan, and San Juan.
75