Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), consisting of the Monitor and test strip, is used for quantitative determination of international normalized ratio (INR) in fresh, capillary whole blood to monitor the effect of warfarin therapy on clotting time by health care professionals. (in vitro diagnostic use).
Enforcement
- Recall Number
- Z-1546-2014
- Event ID
- 68038
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Alere San Diego, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 14, 2014
- Initiation Date
- April 16, 2014
- Classification Date
- May 8, 2014
- Termination Date
- July 19, 2017
- Address
- 9975 Summers Ridge Rd, N/A, San Diego, CA, 92121-2997, United States
Description
Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), consisting of the Monitor and test strip, is used for quantitative determination of international normalized ratio (INR) in fresh, capillary whole blood to monitor the effect of warfarin therapy on clotting time by health care professionals. (in vitro diagnostic use).
Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Lot/Unit Codes: Applicable to all Alere INRatio2 PT/INR Professional Test Strips lots packaged from 8/22/13 through 3/28/14, including the following: 324810, 324811, 324812, 327543, 327544, 327549, 328109, 328705, 328707, 330238, 330852, 331340, 331916, 331922, 332841, 332842, 332843, 332959, 333362, 333654, 333655, 333864, 333865, 334572, 334575, 335552, 335553, 335796, 335797, 336426, 336668, 337350, 338965, 338966, 339257, 339707, 327856V Expiration date, or Expected shelf life: range from 4/30/14 to 1/31/15
US Nationwide Distribution
Estimated 99,795 Total units