FDA Enforcement Class II Ongoing

BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to securely store and dispense medications to a qualified and authorized user in a clinical settings.

Recall: Z-1545-2025 · Reported April 16, 2025

Enforcement

Recall Number
Z-1545-2025
Event ID
96557
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
CareFusion 303, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
April 16, 2025
Initiation Date
March 11, 2025
Classification Date
April 9, 2025
Address
10020 Pacific Mesa Blvd, N/A, San Diego, CA, 92121-4386, United States

Description

BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to securely store and dispense medications to a qualified and authorized user in a clinical settings.

Reason

Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the CUBIE opening more times than needed for the transaction to be completed, which may result in a patient receiving more than the intended dose of medication.

Code Info

REF/UDI-DI(GTIN)/Software Version: 169-143/10885403512568/2.4; 169-145/10885403512568/2.4; 169-144/10885403512568/2.4; 169-142/10885403512568/2.4; 139088-01/1.2.3.23; 155288-01/10885403508721/2.4; 169-137/10885403512544/2.4; 155311-01/10885403510847/2.4; 169-151/10885403512544/1.2.3.23; 138921-01/2.4; 155159-01/10885403504655/2.4; 169-101/10885403512544/2.4; 139122-01/2.4; 155160-01/10885403504662/2.4; 169-102/10885403512544/2.4; 138912-01/2.4; 155161-01/10885403504679/2.4; 169-103/10885403512544/2.4; 138902-01/2.4; 169-104/10885403512544/2.4; 138903-01/2.4; 155163-01/10885403504686/2.4; 169-105/10885403512544/2.4; 138908-01/2.4; 169-106/10885403512544/2.4; 138906-01/2.4; 155165-01/10885403504693/2.4; 169-107/10885403512544/2.4; 169-152/10885403512544/1.2.3.23; 138907-01/2.4; 155166-01/10885403504570/2.4; 169-108/10885403512544/2.4; 139131-01/2.4; 155167-01/10885403504587/2.4; 169-109/10885403512544/2.4; 139091-01/2.4; 155168-01/10885403504594/2.4; 169-110/10885403512544/2.4; 138909-01/2.4; 155169-01/10885403504600/2.4; 169-111/10885403512544/2.4; 138910-01/2.4; 155310-01/10885403510830/2.4; 169-112/10885403512544/2.4; 169-150/10885403512544/1.2.3.23; 138915-01/2.4; 155171-01/10885403504617/2.4; 169-113/10885403512544/2.4; 138943-01/2.4; 155081-01/10885403503870/2.4; 169-23/10885403512568/2.4; 138955-01/2.4; 155082-01/10885403503887/2.4; 169-24/10885403512568/2.4; 138924-01/2.4; 155083-01/10885403503894/2.4; 169-25/10885403512568/2.4; 138960-01/2.4; 155084-01/10885403503900/2.4; 169-26/10885403512568/2.4; 138961-01/2.4; 155085-01/10885403503917/2.4; 169-27/10885403512568/2.4; 138926-01/2.4; 155086-01/10885403503726/2.4; 169-28/10885403512568/2.4; 138927-01/2.4; 155087-01/10885403503733/2.4; 169-29/10885403512568/2.4; 155293-01/10885403509926/2.4; 169-138/10885403512568/2.4; 155303-01/10885403510779/2.4; 139082-01/2.4; 155088-01/10885403503740/2.4; 169-30/10885403512568/2.4; 155305-01/10885403510793/2.4; 139056-01/2.4; 155089-01/10885403503757/2.4; 169-31/10885403512568/2.4; 155304-01/10885403510786/2.4; 139045-01/2.4; 155090-01/10885403503764/2.4; 169-32/10885403512568/2.4; 138932-01/2.4; 155091-01/10885403503771/2.4; 169-33/10885403512568/2.4; 139084-01/2.4; 155092-01/10885403503788/2.4; 169-34/10885403512568/2.4; 138963-01/2.4; 155093-01/10885403503795/2.4; 169-35/10885403512568/2.4; 155294-01/10885403509933/2.4; 169-139/10885403512568/2.4; 138938-01/2.4; 155094-01/10885403503801/2.4; 169-36/10885403512568/2.4; 139125-01/2.4; 155095-01/10885403503818/2.4; 169-37/10885403512568/2.4; 138919-01/2.4; 155096-01/10885403503696/2.4; 169-38/10885403512568/2.4; 139126-01/2.4; 155097-01/10885403504198/2.4; 169-39/10885403512568/2.4; 139127-01/2.4; 155098-01/10885403503702/2.4; 169-40/10885403512568/2.4; 139043-01/2.4; 155099-01/10885403504204/2.4; 169-41/10885403512568/2.4; 138933-01/2.4; 155100-01/10885403504211/2.4; 169-42/10885403512568/2.4; 155299-01/10885403510526/2.4; 169-140/10885403512568/1.2.3.23; 139142-01/2.4; 155101-01/10885403504228/2.4; 169-43/10885403512568/2.4; 155302-01/10885403510762/2.4; 138959-01/2.4; 155102-01/10885403504235/2.4; 169-44/10885403512568/2.4; 138954-01/2.4; 155103-01/10885403504242/2.4; 169-45/10885403512568/2.4; 138934-01/2.4; 155104-01/10885403504259/2.4; 169-46/10885403512568/2.4; 138950-01/2.4; 155105-01/10885403503719/2.4; 169-47/10885403512568/2.4; 138947-01/2.4; 155106-01/10885403504266/2.4; 169-48/10885403512568/2.4; 138945-01/2.4; 155107-01/10885403504273/2.4; 169-49/10885403512568/2.4; 138931-01/2.4; 155108-01/10885403504280/2.4; 169-50/10885403512568/2.4; 138929-01/2.4; 155109-01/10885403504327/2.4; 169-51/10885403512568/2.4; 138946-01/2.4; 155110-01/10885403504297/2.4; 169-52/10885403512568/2.4; 138940-01/2.4; 155111-01/10885403504303/2.4; 169-53/10885403512568/2.4; 138952-01/2.4; 155112-01/10885403504099/2.4; 169-54/10885403512568/2.4; 138917-01/2.4; 155113-01/10885403504105/2.4; 169-55/10885403512568/2.4; 138918-01/2.4; 155114-01/10885403504112/2.4; 169-56/10885403512568/2.4; 138941-01/2.4; 155115-01/10885403504129/2.4; 169-57/10885403512568/2.4; 138905-01/2.4; 155116-01/10885403504136/2.4; 169-58/10885403512568/2.4; 138916-01/2.4; 155117-01/10885403504143/2.4; 169-59/10885403512568/2.4; 139141-01/2.4; 155118-01/10885403504150/2.4; 169-60/10885403512568/2.4; 138914-01/2.4; 155119-01/10885403504334/2.4; 169-61/10885403512568/2.4; 139063-01/2.4; 155120-01/10885403504167/2.4; 169-62/10885403512568/2.4; 138953-01/2.4; 155121-01/10885403504174/2.4; 169-63/10885403512568/2.4; 138936-01/2.4; 155122-01/10885403504181/2.4; 169-64/10885403512568/2.4; 138962-01/2.4; 155123-01/10885403504082/2.4; 169-65/10885403512568/2.4; 138944-01/2.4; 155124-01/10885403504075/2.4; 169-66/10885403512568/2.4; 138957-01/2.4; 155125-01/10885403503658/2.4; 169-67/10885403512568/2.4; 138922-01/2.4; 155126-01/10885403503665/2.4; 169-68/10885403512568/2.4; 138930-01/2.4; 155127-01/10885403503672/2.4; 169-69/10885403512568/2.4; 139128-01/2.4; 155128-01/10885403503689/2.4; 169-70/10885403512568/2.4; 138958-01/2.4; 155129-01/10885403504341/2.4; 169-71/10885403512568/2.4; 138920-01/2.4; 155130-01/10885403504358/2.4; 169-72/10885403512568/2.4; 138937-01/2.4; 155131-01/10885403504365/2.4; 169-73/10885403512568/2.4; 139086-01/2.4; 155132-01/10885403504372/2.4; 169-74/10885403512568/2.4; 138942-01/2.4; 155133-01/10885403503542/2.4; 169-75/10885403512568/2.4; 138949-01/2.4; 155134-01/10885403503559/2.4; 169-76/10885403512568/2.4; 139062-01/2.4; 155135-01/10885403503566/2.4; 169-77/10885403512568/2.4; 138939-01/2.4; 155136-01/10885403503573/2.4; 169-78/10885403512568/2.4; 138948-01/2.4; 155137-01/10885403503597/2.4; 169-79/10885403512568/2.4; 138956-01/2.4; 155138-01/10885403503603/2.4; 169-80/10885403512568/2.4; 138923-01/2.4; 155139-01/10885403503610/2.4; 169-81/10885403512568/2.4; 139163-01/2.4. 1119-00/10885403512704/2.4; 1139-00/10885403512704/2.4. 139054-01/2.4; 139055-01/2.4; 139053-01/2.4; 139041-01/2.4; 139046-01/2.4; 139038-01/2.4; 139040-01/2.4; 139049-01/2.4; 139044-01/2.4; 139039-01/2.4; 139051-01/2.4; 139164-01/2.4; 139052-01/2.4; 139070-01/2.4; 139071-01/2.4; 139061-01/2.4; 139066-01/2.4; 139065-01/2.4; 139058-01/2.4; 139060-01/2.4; 139067-01/2.4; 139064-01/2.4; 139059-01/2.4; 139068-01/2.4; 139165-01/2.4; 139069-01/2.4. 1144-00/2.4; 1137-00/2.4; 1147-00/10885403517181/2.4 1145-00/10885403517167/2.4

Distribution

US Nationwide distribution: CA, TX, WI, CO, AK, VA, TN, WA, NC, NY, MS, FL, KS, MI, IN, MN, MT, KY, MO, UT, NJ, OH, PA, ND, OR, NE, SD, NV, OK, GA, MA, IA, HI, ME, AZ, IL, SC, LA, WV, AL, ID, NM, DC, NH, VT, CT, MD, AR, WY.

Quantity

7357