FDA Enforcement Class II Terminated

12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.

Recall: Z-1544-2014 · Reported May 14, 2014

Enforcement

Recall Number
Z-1544-2014
Event ID
68045
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 14, 2014
Initiation Date
April 16, 2014
Classification Date
May 2, 2014
Termination Date
August 18, 2014
Address
5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States

Description

12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.

Reason

Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.

Code Info

Model number: 0242000012; Lot/serial numbers: 021464-2, 39824-1-1, 4014, 643893-1-4, 27419, 42901-1-1, 35728-2-1, 48639-1-8, 021464-3, 40288-1-1, 43020, 43893-1-6, 27701, 43104-1-1, 36175-1-1, 48956-1-1, 23027, 40629-3-,1, 43622, 43893-1-7, 29129, 43270-1-1, 36421-1-1, 48956-1-1A, 23398, 40721-1-1, 50476, 43893-1-8, 29670, 43523-1-1, 36424-1-1, 51304-1-1, 23542, 41231-4-1, 54990, 43893-1-9, 30250, 43523-1-1A, 37251-1-1, 51304-1-1A, 23846, 41478-1-1, 7J1, 48639-1-2, 30919, 43674-1-1, 37392-1-1, 52123-1-1, 24073, 41745-1-1, 123, 48639-1-3, 3218., 43893-1-1, 37794-1-1, 52956-1-1, 24511, 42086-1-1, 28646-1-1, 48639-1-4, 33226, 43893-1-10, 38867-1-1, 54346-1-1, 26445, 42306-1-1, 29005-1-1, 48639-1-5, 33961, 43893-1-1, 1SS15327, 26843, 42526-1-1, 30384-1-1, 48639-1-7, 35933, 43893-1-12, 37991, 36610, 43893-1-13, 39219, 43893-1-2, 43893-1-14.

Distribution

Worldwide distribution

Quantity

2591 trays