FDA Enforcement
Class II
Terminated
RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763 A ( 3 motors, autocontour) 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
Recall: Z-1541-2014
·
Reported May 14, 2014
Enforcement
- Recall Number
- Z-1541-2014
- Event ID
- 67871
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Med-Mizer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 14, 2014
- Initiation Date
- April 18, 2014
- Classification Date
- May 2, 2014
- Termination Date
- May 23, 2016
- Address
- 80 Commerce Dr, N/A, Batesville, IN, 47006-6700, United States
Description
RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763 A ( 3 motors, autocontour) 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.
Reason
All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on th
Code Info
Model 762: 300000-300191 Model 763: 200000 Model 763A: 400000-40103
Distribution
USA (nationwide)
Quantity
191