9.6F Plastic Dignity Mid-Sized CT Port w/Pre-Attached Silicone Catheter The CT Power Injectable Implantable Infusion Ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Power Injectable Implantable Infusion Port device is indicated for power injection of contrast media
Enforcement
- Recall Number
- Z-1536-2017
- Event ID
- 76625
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medical Components, Inc dba MedComp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 29, 2017
- Initiation Date
- February 23, 2017
- Classification Date
- March 22, 2017
- Termination Date
- February 2, 2018
- Address
- 1499 Delp Dr, Harleysville, PA, 19438-2936, United States
Description
9.6F Plastic Dignity Mid-Sized CT Port w/Pre-Attached Silicone Catheter The CT Power Injectable Implantable Infusion Ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Power Injectable Implantable Infusion Port device is indicated for power injection of contrast media
The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit contains a 10F valved peelable introducer. The kit is packaged with a 9F valved peelable introducer. The port lumen will not fit through the introducer during the insertion procedure.
MRCTI96801 Lot# MHWQ060 S2 , UDI# 884908031232
Worldwide Distribution to PR and Panama
71 units