90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, intertrochanteric valgus osteotomies, proximal and distal tibial osteotomies and humeral fractures and osteotomies.
Enforcement
- Recall Number
- Z-1536-2013
- Event ID
- 65091
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- OrthoPediatrics Corp
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- June 19, 2013
- Initiation Date
- April 19, 2013
- Classification Date
- June 12, 2013
- Termination Date
- September 19, 2013
- Address
- 2850 Frontier Dr, N/A, Warsaw, IN, 46582-7001, United States
Description
90 Degree Cannulated Infant Blade Plate 35mm x 5mm x 3 Hole Product Usage: The OrthoPediatrics Locking Cannulated Blade Plate System is intended for fixation of long bone fractures and osteotomies in all pediatric subgroups (except neonates) and in small stature adults. Specific indications include: intertrochanteric derotation and varus osteotomies, femoral neck and pertrochanteric fractures, intertrochanteric valgus osteotomies, proximal and distal tibial osteotomies and humeral fractures and osteotomies.
Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.
All lot numbers are affected. Catalog No. 00-1200-1002
USA Nationwide Distribution
130 devices were shipped