FDA Enforcement Class I Ongoing

AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.

Recall: Z-1535-2024 · Reported April 24, 2024

Enforcement

Recall Number
Z-1535-2024
Event ID
94302
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
SonarMed Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 24, 2024
Initiation Date
March 25, 2024
Classification Date
April 18, 2024
Address
12220 N Meridian St Ste 150, Carmel, IN, 46032-6972, United States

Description

AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.

Reason

Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.

Code Info

Model No. AW-M0001; GTIN: 00851334007001; Serial No. AW0572, AW0573, AW0574, AW0577, AW0578, AW0579, AW0580, AW0581, AW0584, AW0585, AW0586, AW0588, AW0591, AW0593, AW0594, AW0600, AW0606, AW0607, AW0615, AW0616, AW0624, AW0626, AW0627, AW0648, AW0650, AW0651, AW0658, AW0664, AW0672, AW0674, AW0675, AW0678, AW0680, AW0681, AW0684, AW0685, AW0701, AW0702, AW0703, AW0704, AW0705, AW0706, AW0707, AW0708, AW0709, AW0710, AW0711, AW0712, AW0718, AW0719, AW0720, AW0721, AW0722, AW0723, AW0724, AW0725, AW0726, AW0727, AW0728, AW0729, AW0730, AW0731, AW0732, AW0733, AW0734, AW0735, AW0736, AW0737, AW0738, AW0739, AW0740, AW0741, AW0742, AW0743, AW0744, AW0745, AW0746, AW0747, AW0748, AW0749, AW0763, AW0764, AW0765, AW0766, AW0767, AW0768, AW0769, AW0770, AW0771, AW0772, AW0773, AW0774, AW0775, AW0776, AW0777, AW0778, AW0779, AW0780, AW0781, AW0782, AW0783, AW0784, AW0785, AW0786, AW0787, AW0788, AW0789, AW0790, AW0791, AW0792, AW0793, AW0794, AW0795, AW0796, AW0797, AW0798, AW0799, AW0800, AW0804, AW0805, AW0814, AW0817, AW0818, AW0819, AW0820, AW0821, AW0825, AW0826, AW0827, AW0828, AW0830, AW0831, AW0832, AW0833, AW0834, AW0835, AW0836, AW0837, AW0838, AW0839, AW0840, AW0841, AW0842, AW0854, AW0855.

Distribution

_AK, AZ, CA, FL, IL, KY, MD, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, & WA_

Quantity

145 units