IMMULITE 2000/2000 XPi Systems Androstenedione (Test Code AND), REF L2KAO2, SMN 10381188, IVD, 200 test kit size --- Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL55 4EL, UK For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and 2000 Systems Analyzers for the quantitative measurement of 4-androstenedione in human serum.
Enforcement
- Recall Number
- Z-1532-2014
- Event ID
- 67211
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 7, 2014
- Initiation Date
- December 20, 2013
- Classification Date
- April 30, 2014
- Termination Date
- February 28, 2017
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005, United States
Description
IMMULITE 2000/2000 XPi Systems Androstenedione (Test Code AND), REF L2KAO2, SMN 10381188, IVD, 200 test kit size --- Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL55 4EL, UK For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and 2000 Systems Analyzers for the quantitative measurement of 4-androstenedione in human serum.
Siemens Healthcare Diagnostics confirmed over-recovery of samples with concentrations greater than 5.5 ng/mL with the IMMULITE/IMMULITE 1000 and IMMULITE 2000/IMMULITE 2000 XPi Androstenedione assays (LKAO1, L2KAO2). Siemens has determined that these lots are linear up to a concentration of 5.5 ng/mL, instead of 10 ng/mL as stated in the Instructions for Use (IFU). Quality Control materials at concentrations between 5.5 and 10 ng/mL will detect this issue.
Lot Numbers 314 and above
Worldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, China (People's Republic of), Czech.Rebuplik, Denmark, Estonia, France, Germany, Great Britain, Italy, New Zealand, Portugal, Spain, Switzerland, Turkey and Vatican City.
7741