FDA Enforcement Class II Ongoing

GEM Premier 5000 PAK; Part No. 00055360010.

Recall: Z-1526-2026 · Reported March 18, 2026

Enforcement

Recall Number
Z-1526-2026
Event ID
98349
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Instrumentation Laboratory
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
March 18, 2026
Initiation Date
January 20, 2026
Classification Date
March 10, 2026
Address
180 Hartwell Rd, N/A, Bedford, MA, 01730-2443, United States

Description

GEM Premier 5000 PAK; Part No. 00055360010.

Reason

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Code Info

Part number: 00055360010; UDI: 08426950807599; Lots No.: 251023N, 251024I, 251103I, 251117T, 251118B, 251120R, 251124O, 251126E, 251201X, 251202E, 251202G, 251203H, 251204O, 251205L, 251205N, 251208C, 251208G, 251209Q, 251209R, 251215Y, 251217V, 251218AA, 251218T, 251219M, 251222AE, 251222AF, 251222AG, 251222E, 251222F, 251223R, 251223S, 251229H, 251229J, 251229U, 251230A, 251231E, 251231K, 251231L.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam.

Quantity

1,225 units