Implantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Azure family: Azure XT DR MRI SureScan", Azure S DR MRI SureScan, Azure XT SR MRI SureScan, Azure S SR MRI SureScan. 2. Astra family: Astra XT DR MRI SureScan, Astra XT SR MRI SureScan.
Recall: Z-1525-2021
·
Reported May 12, 2021
Enforcement
- Recall Number
- Z-1525-2021
- Event ID
- 87662
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 12, 2021
- Initiation Date
- April 6, 2021
- Classification Date
- May 4, 2021
- Address
- 8200 Coral Sea St Ne, N/A, Mounds View, MN, 55112-4391, United States
Description
Implantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure. 1. Azure family: Azure XT DR MRI SureScan", Azure S DR MRI SureScan, Azure XT SR MRI SureScan, Azure S SR MRI SureScan. 2. Astra family: Astra XT DR MRI SureScan, Astra XT SR MRI SureScan.
Reason
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
Code Info
all codes
Distribution
Worldwide distribution.
Quantity
575815 units