Medtronic Activa RC (37612), Activa PC (37601), Activa SC (37602), Activa SC (37603), Activa PC+S (37604), Multi-program Rechargeable Neurostimulator for Deep Brain Stimulation for movement disorders.
Enforcement
- Recall Number
- Z-1525-2014
- Event ID
- 66841
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medtronic Neuromodulation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 7, 2014
- Initiation Date
- February 27, 2014
- Classification Date
- April 29, 2014
- Termination Date
- November 7, 2014
- Address
- 7000 Central Ave NE, Minneapolis, MN, 55432-3568, United States
Description
Medtronic Activa RC (37612), Activa PC (37601), Activa SC (37602), Activa SC (37603), Activa PC+S (37604), Multi-program Rechargeable Neurostimulator for Deep Brain Stimulation for movement disorders.
Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval
The first released affected product was released to the market in April, 2005.
Nationwide Distribution - all states including DC and Puerto Rico.
170,286 US, 73,921 OUS for all devices