FDA Enforcement Class II Terminated

Cervical Spine Locking Plate (CSLP) System, 4mm Cancellous Self Tapping Screw, Length 16mm Lot # 9666776 and 4mm Cortex Standard Screw, Length 16mm Lot # 9671879

Recall: Z-1524-2016 · Reported April 27, 2016

Enforcement

Recall Number
Z-1524-2016
Event ID
73549
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes (USA) Products LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 27, 2016
Initiation Date
January 6, 2016
Classification Date
April 19, 2016
Termination Date
January 31, 2017
Address
1301 Goshen Pkwy, N/A, West Chester, PA, 19380-5986, United States

Description

Cervical Spine Locking Plate (CSLP) System, 4mm Cancellous Self Tapping Screw, Length 16mm Lot # 9666776 and 4mm Cortex Standard Screw, Length 16mm Lot # 9671879

Reason

DePuy Synthes is initiating a Voluntary Medical Device Recall of lot # 9666776 and 9671879 of Cancellous and Cortex 4.0 mm Screws, which are a part of the Cervical Spine Locking Plate (CSLP) System due to discovery of a single lot of 450.136 screws (4mm Cancellous Self Tapping Screw, Length 16mm, lot 9666776) being swapped with a single lot of 450.016 (4mm Cortex Standard Screw, length 16 mm, lot 9671879).

Code Info

Lot # 9666776 (catalog ID 450.136) and 9671879 (catalog ID 450.016)

Distribution

Worldwide Distribution - US Nationwide in the states of NC, CA, NY, TX, NJ, CO and the countries of: Ontario and Canada.

Quantity

30 units of Lot # 9666776 and 11 units of Lot # 9671879