FDA Enforcement Class II Terminated

MicroScan Pos Combo Panel Type 33 (PC33): Catalog number B1017-211 MicroScan Pos Combo 33 B1017-211 panels are designed for use in determining quantitative and /or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive bacteria.

Recall: Z-1524-2015 · Reported May 6, 2015

Enforcement

Recall Number
Z-1524-2015
Event ID
70904
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 6, 2015
Initiation Date
April 8, 2014
Classification Date
April 27, 2015
Termination Date
April 27, 2015
Address
2040 Enterprise Blvd, West Sacramento, CA, 95691-3427, United States

Description

MicroScan Pos Combo Panel Type 33 (PC33): Catalog number B1017-211 MicroScan Pos Combo 33 B1017-211 panels are designed for use in determining quantitative and /or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive bacteria.

Reason

Siemens investigation confirmed that a portion one lot of panel Pos Combo 33 lot was packaged without desiccants. The substrates on a panel packaged without a desiccant may deteriorate over time due to exposure to additional moisture in the panel pouch.

Code Info

Catalog number B10170211; Siemens Material number (SMN): 1071530, lot number 2014-09-12.: Expiration date 2014-09-12.

Distribution

Nationwide Distribution including Puerto Rico and to the states of : CT, GA, NC, OK, NY, SC, KY, TN, CA, WA, OR, NJ, OH, PA, FL, MI, TX, WI, MO and NH.

Quantity

664 panel boxes