FDA Enforcement Class II Terminated

Power electronic box, Part #AS-0066, subassembly for the MyoCycle MC-2.

Recall: Z-1523-2022 · Reported August 24, 2022

Enforcement

Recall Number
Z-1523-2022
Event ID
90620
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Myolyn Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 24, 2022
Initiation Date
July 19, 2022
Classification Date
August 16, 2022
Termination Date
September 23, 2024
Address
6931 Nw 22nd St Ste A, Gainesville, FL, 32653-1231, United States

Description

Power electronic box, Part #AS-0066, subassembly for the MyoCycle MC-2.

Reason

Screws mounting an internal power supply may come loose during shipping resulting in increased risk of high voltage electrical shock.

Code Info

Serial numbers 000800 thru 000819, 000821 thru 000822, 000824 thru 000827, 000829 thru 000831, 000835, 000836, 000840 thru 000842, 000850 thru 000852, and 000857 thru 000859; UDI 000861553000417

Distribution

Distribution was made to AZ, CA, CT, FL, GA, ID, IL, KY, MA, MI, MN, MO, MN, NC, NJ, NM, NY, OH, OR, TX, UT, VA, WA, and WI. There was no foreign/military/government distribution.

Quantity

45 devices