FDA Enforcement
Class II
Terminated
Anatomical Shoulder Handle for Rasp Item: 01.04233.000 Used during total joint arthroplasty of the shoulder.
Recall: Z-1523-2015
·
Reported May 6, 2015
Enforcement
- Recall Number
- Z-1523-2015
- Event ID
- 70993
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Gmbh
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 6, 2015
- Initiation Date
- March 18, 2015
- Classification Date
- April 27, 2015
- Termination Date
- September 23, 2015
- Address
- Sulzer Allee 8, Winterthur, N/A, Switzerland
Description
Anatomical Shoulder Handle for Rasp Item: 01.04233.000 Used during total joint arthroplasty of the shoulder.
Reason
Complaints of difficulties to attach the Anatomical Shoulder Rasps (high resistance) or, once attached, due to difficulties to remove the handle (seizing up of the two components). This has the potential for delay in surgery.
Code Info
Lot Number(s): 13870129 & 13878179.
Distribution
Distributed to the states of IN, GA, WI, MO, PA, KS, TN, and NY.
Quantity
30 units