FDA Enforcement Class I Terminated

Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons in cardiovascular procedures.

Recall: Z-1523-2014 · Reported May 28, 2014

Enforcement

Recall Number
Z-1523-2014
Event ID
67815
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Maquet Datascope Corp - Cardiac Assist Division
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 28, 2014
Initiation Date
March 21, 2014
Classification Date
May 21, 2014
Termination Date
March 20, 2018
Address
1300 MacArthur Blvd., N/A, Mahwah, NJ, 07430-2052, United States

Description

Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons in cardiovascular procedures.

Reason

Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.

Code Info

IABP Model Numbers System 98 0998-00-0446-xx 0998-UC-0446-xx System 98XT 0998-00-0479-xx 0998-UC-0479-xx CS100i 0998-UC-0446Hxx 0998-UC-0479Hxx CS100 0998-00-3013-xx 0998-UC-3013-xx CS300 0998-00-3023-xx 0998-UC-3023-xx

Distribution

Worldwide Distribution - USA (nationwide) and approximately 104 countries.

Quantity

11,881 units (US 5183 and OUS 6698)