FDA Enforcement Class II Ongoing

Endo-Model Replacement Plateau; Item Number: 15-0027/11;

Recall: Z-1521-2026 · Reported March 18, 2026

Enforcement

Recall Number
Z-1521-2026
Event ID
98342
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 18, 2026
Initiation Date
January 12, 2026
Classification Date
March 10, 2026
Address
Ost Str. 4-10, N/A, Norderstedt, N/A, N/A, Germany

Description

Endo-Model Replacement Plateau; Item Number: 15-0027/11;

Reason

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Code Info

Item Number: 15-0027/11; UDI-DI: 04026575431953; Serial/Lot Number: 2549126, 2549127;

Distribution

Worldwide - US Nationwide distribution in the state of NJ and the countries of Belarus, Canada, Switzerland , Colombia, Germany, Spain, United Kingdom, Hungary, India, Italy, Netherlands, Romania, Saudi Arabia.

Quantity

2 units (OUS only)