FDA Enforcement Class II Terminated

Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test

Recall: Z-1519-2018 · Reported May 2, 2018

Enforcement

Recall Number
Z-1519-2018
Event ID
79526
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Diagnostics Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 2, 2018
Initiation Date
January 2, 2018
Classification Date
April 25, 2018
Termination Date
April 25, 2019
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test

Reason

The previous version of method sheets for the Acetaminophen assay on the COBAS INTEGRA 400 plus/800/800 CTS analyzers and the MODULAR ANALYTICS P module are missing the following statement in the Limitations-interference section: "In very rare cases, gammopathy, in particular type IgM (Waldenstr¿m s macroglobulinemia), may cause unreliable results."

Code Info

Lot codes: 11021500 7/31/2016 12361201 5/31/2017 11031700 9/30/2018 15247301 8/31/2017 11071400 1/31/2016 18036001 11/30/2017 11071600 8/10/2016 19329601 1/31/2018 11101500 1/21/2016 20511301 5/31/2018 11111600 4/30/2018 24076801 8/31/2018 60482201 2/29/2016 60852601 5/31/2016 61385701 9/30/2016 62032101 1/31/2017

Distribution

Nationwide Distribution

Quantity

35,111