Light Sheer Desire Diode Laser System with XC Handpiece Accessory options.
Enforcement
- Recall Number
- Z-1519-2015
- Event ID
- 70863
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Lumenis Limited
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 6, 2015
- Initiation Date
- March 25, 2015
- Classification Date
- April 24, 2015
- Termination Date
- June 15, 2016
- Address
- 13 Hayetzira St.,Yokneam Ind. Park, Yokneam, N/A, Israel
Description
Light Sheer Desire Diode Laser System with XC Handpiece Accessory options.
Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended settings. Operator Manual parameters are lower than indicated for specific hair color and Fitzpatrick skin type resulting in insufficient treatment effect. May result in patient burns and hypopigmentation.
Lot numbers: XI0017, XI0018, XI0023, XIP0043, 182, 184, 185, XIP0049, 236, 237, 239, 240, 242, 248, 249, 250, 251, 254, 256, 257, XIP0038, XIP0131, XIP0125, 3048, 3052, 3065, 3074, XI0012, XIP0143, XIP0144, XI0001, XIP0072, XI0028, XIP0061, XI0028, XI0036, XIP0074, XIP0077, XIP0082, XIP0087, XIP0132, XIP0127, XIP0124, XIP0129, XI0008, XI0029, XI0037, XI0046, XIP0123, XIP0059, XIP0107, XIP0089, XIP0096, XIP0156, XI0005, XIP0114, XI0014, XI0022, XI0015, XI0021, XI0031, XI0034, XI0024, XI0026, XIP0047, XI0042, XI0039, XIP0044, XIP0025, XIP0035, XIP0041, XIP0030, XIP0051, XIP0057, XIP0113, XIP0067, XIP0053, XIP0083, XIP0058, XIP0055, XIP0060, XIP0066, XIP0163, XIP0071, XIP0063, XIP0120, XIP0088, XIP0081, XIP0064, XIP0050, XIP0101, XIP0062, XIP0095, XIP0121, XIP0094, XIP0108, XIP0105, XIP0122, XIP0119, XIP0104, XIP0097, 104, XIP0065, XIP0164, XIP0159, XIP0152, XIP0166, XI0016, XIP0045, XIP0052, XIP0158, XIP0092, XIP0111, XIP0086, XIP0086, XIP0086, XIP0112, XIP0130, XIP0157, XIP0093, XIP0151, XIP0149, XIP0167, XIP0168, XIP0162, XIP0171, XIP0138, XIP0118
Worldwide distribution. US nationwide (to OR, NM, NY, IL, and CO), Switzerland, Canada, Italy, Great Britain, United Arab Emirates, France, Belgium, India, Israel, Japan, Portugal, Czech Republic, Kuwait, Germany, Spain, Georgia, Armenia, Slovenia, New Zealand, Saudi Arabia, and Russian Federation.
117 units