FDA Enforcement Class II Terminated

Maquet Getinge-BEQ-TOP 15705 Material: 701051122

Recall: Z-1514-2019 · Reported June 5, 2019

Enforcement

Recall Number
Z-1514-2019
Event ID
82703
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Maquet Cardiovascular, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 5, 2019
Initiation Date
April 18, 2019
Classification Date
May 30, 2019
Termination Date
July 28, 2025
Address
45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States

Description

Maquet Getinge-BEQ-TOP 15705 Material: 701051122

Reason

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Code Info

Batch Numbers: 3000044566 3000046394 3000049202 3000054550 3000058043 3000060239 3000066276 3000069368 3000071833 3000074339 3000080701 3000081598

Distribution

Nationwide

Quantity

N/A