FDA Enforcement Class II Terminated

Stem Extractor f/Guide Bar, Part number 03.401.072. Instrumentation for extraction and revision of Epoca shoulder implants.

Recall: Z-1514-2018 · Reported May 2, 2018

Enforcement

Recall Number
Z-1514-2018
Event ID
79698
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 2, 2018
Initiation Date
March 19, 2018
Classification Date
April 25, 2018
Termination Date
April 11, 2019
Address
1302 Wrights Ln E, N/A, West Chester, PA, 19380-3417, United States

Description

Stem Extractor f/Guide Bar, Part number 03.401.072. Instrumentation for extraction and revision of Epoca shoulder implants.

Reason

There is a potential for the connection screw of the stem extractor to break.

Code Info

09-6793, 10-2637, 10-5070, 12-2707, 14-1085, 15-8949

Distribution

Worldwide Distribution: US (nationwide) and to countries of: Canada and Switzerland.

Quantity

31