FDA Enforcement Class II Ongoing

Xhibit Telemetry Receiver, Model: 96280

Recall: Z-1513-2022 · Reported August 17, 2022

Enforcement

Recall Number
Z-1513-2022
Event ID
90529
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Spacelabs Healthcare, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 17, 2022
Initiation Date
May 31, 2022
Classification Date
August 11, 2022
Address
35301 Se Center St, Snoqualmie, WA, 98065-9216, United States

Description

Xhibit Telemetry Receiver, Model: 96280

Reason

Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.

Code Info

UDI-DI: 10841522100246, Software version 1.4.0. Serial Numbers: 6280-001538, 6280-000590, 6280-000589, 6280-000339, 9280-000711, 6280-001103, 6280-001102, 6280-001469, 6280-001467, 6280-001686, 6280-001399, 6280-001323, 6280-000457, 6280-000456, 6280-000455, 6280-000440, 6280-000439, 6280-000438, 6280-000429, 6280-000476, 6280-000475, 6280-001693, 6280-000703, 6280-000702, 6280-000701, 6280-000700, 6280-000314, 6280-000258, 6280-000157, 6280-001223, 6280-001222, 6280-001221, 6280-001220, 6280-001219, 6280-001218, 6280-001217, 6280-001216, 6280-001215, 6280-001214, 6280-001207, 6280-001127, 6280-001125, 6280-001124, 6280-001123, 6280-001056, 6280-001055, 6280-001054, 6280-001053, 6280-001052, 6280-001051, 6280-001050, 6280-001288, 6280-001287, 6280-001490, 6280-001692, 6280-000588, 6280-000517, 6280-000516, 6280-000515, 6280-001668, 6280-000206, 6280-000162, 6280-000210, 6280-000663, 6280-000377, 6280-001589, 6280-001385, 6280-001384, 6280-001382, 6280-001405, 6280-000594, 6280-000593, 6280-000592, 6280-000596, 6280-000274, 6280-001844, 6280-001843, 6280-001721, 6280-001720, 6280-000136, 6280-001784, 6280-001783, 6280-000359, 6280-000340, 6280-000336, 6280-001718, 6280-001706, 6280-001705, 6280-001704, 6280-001703, 6280-001702, 6280-001699, 6280-001698, 6280-001672, 6280-001669, 6280-000622, 6280-001797, 6280-000097, 6280-001558, 6280-001551, 6280-001649, 6280-001648, 6280-001647, 6280-001646, 6280-001645, 6280-001594, 6280-000093, 6280-001168, 6280-001531, 6280-001670, 6280-000219, 6280-000550, 6280-000878, 6280-000877, 6280-000872, 6280-000862, 6280-000858, 6280-000857, 6280-000221, 6280-000089, 6280-000204, 6280-000212, 6280-000510, 6280-000509, 6280-000209, 6280-000320, 6280-000538, 6280-000528, 6280-000205, 6280-000361, 6280-000360, 6280-000223, 6280-000963, 6280-000529, 6280-001575, 6280-001574, 6280-001573, 6280-001572, 6280-001571, 6280-001570, 6280-000467, 6280-000466, 6280-000465, 6280-000464, 6280-000463, 6280-000462, 6280-000461, 6280-000460, 6280-000459, 6280-000458, 6280-000222, 6280-000264, 6280-000275, 6280-000354, 6280-001677, 6280-001048, 6280-001043, 6280-001041, 6280-001038, 6280-001037, 6280-001036, 6280-001023, 6280-001022, 6280-001016, 6280-000787, 6280-001418, 6280-001419, 6280-001416, 6280-000894, 6280-000866, 6280-000865, 6280-000863, 6280-000902, 6280-000898, 6280-000724, 6280-000420, 6280-000419, 6280-000798, 6280-000325, 6280-000532, 6280-000521, 6280-000213, 6280-000254, 6280-000856, 6280-000855, 6280-000854, 6280-000853, 6280-000852, 6280-000851, 6280-001143, 6280-001142, 6280-001141, 6280-000472, 6280-000375, 6280-000368, 6280-000421, 6280-000519, 6280-001213, 6280-000836, 6280-000835, 6280-000834, 6280-000833, 6280-000832, 6280-000831, 6280-000586, 6280-001473, 6280-000319, 6280-000263, 6280-000312, 6280-000579, 6280-000092, 6280-000090, 6280-000084, 6280-000083, 6280-000079, 6280-001380, 6280-001466, 6280-001458, 6280-001611, 6280-001592, 6280-001591, 6280-000388, 6280-000161, 6280-000160, 6280-000159, 6280-000158, 6280-000066, 6280-000053, 6280-000174, 6280-000173, 6280-000172, 6280-000170, 6280-000168, 6280-000167, 6280-000077, 6280-000598, 6280-000355, 6280-000344, 6280-000341, 6280-000338, 6280-000468, 6280-000423, 6280-000478, 6280-000477, 6280-000454, 6280-000428, 6280-000261, 6280-001614, 6280-001642, 6280-001641, 6280-001640, 6280-001639, 6280-001638, 6280-001637, 6280-001636, 6280-001635, 6280-001634, 6280-001633, 6280-001632, 6280-001631, 6280-001630, 6280-001629, 6280-001913, 6280-000138, 6280-000437, 6280-000436, 6280-000177, 6280-000176, 6280-000337, 6280-001857, 6280-001853, 6280-001749, 6280-001359, 6280-001160, 6280-001159, 6280-001158, 6280-001157, 6280-001156, 6280-001154, 6280-001132, 6280-001112, 6280-001109, 6280-000991, 6280-000983, 6280-001189, 6280-001188, 6280-001205, 6280-001204, 6280-001203, 6280-001202, 6280-001200, 6280-001194, 6280-001193, 6280-001184, 6280-001166, 6280-000358, 6280-000357, 6280-000345, 6280-000432, 6280-000431, 6280-000811, 6280-000693, 6280-000690, 6280-000929, 6280-001063, 6280-001062, 6280-001061, 6280-001060, 6280-000391, 6280-001273, 6280-001554, 6280-001553, 6280-001261, 6280-000426, 6280-001525, 6280-000229, 6280-000876, 6280-000875, 6280-000745, 6280-000208, 6280-000574, 6280-000218, 6280-001087, 6280-000045, 6280-000041, 6280-000374, 6280-000373, 6280-001257, 6280-001256, 6280-001255, 6280-001254, 6280-001253, 6280-001252, 6280-001251, 6280-001250, 6280-001249, 6280-000840, 6280-000486, 6280-001825, 6280-000543, 6280-000542, 6280-000839, 6280-000879, 6280-001325, 6280-001071, 6280-001070, 6280-001069, 6280-000900, 6280-000897, 6280-001822, 6280-001874, 6280-001872, 6280-001871, 6280-000695, 6280-000540, 6280-000541, 6280-000539, 6280-000166, 6280-000624, 6280-000614, 6280-000613, 6280-000612, 6280-000611, 6280-000610, 6280-000609, 6280-000608, 6280-000607, 6280-000630, 6280-000629, 6280-000628, 6280-000627, 6280-000626, 6280-001644, 6280-001619, 6280-001617, 6280-000587, 6280-000490, 6280-000489, 6280-000488, 6280-000487, 6280-000974, 6280-000973, 6280-000972, 6280-000971, 6280-000970, 6280-000969, 6280-000968, 6280-000967, 6280-000966, 6280-000965, 6280-000329, 6280-000326, 6280-000318, 6280-001621, 6280-001609, 6280-001596, 6280-001536, 6280-001381, 6280-001340, 6280-001339, 6280-001338, 6280-001337, 6280-001336, 6280-001335, 6280-001334, 6280-001315, 6280-001312, 6280-001311, 6280-000547, 6280-000546, 6280-000545, 6280-000544, 6280-000518, 6280-001764, 6280-001752, 6280-001751, 6280-001746, 6280-001548, 6280-001408, 6280-001431, 6280-001729, 6280-001046, 6280-000601, 6280-001350, 6280-000689, 6280-001487, 6280-001475, 6280-000928, 6280-000927, 6280-000926, 6280-000925, 6280-000901, 6280-001329, 6280-001331, 6280-001330, 6280-001342, 6280-001341, 6280-000514, 6280-000513, 6280-000512, 6280-000511, 6280-000530, 6280-000108, 6280-000107, 6280-000106, 6280-000105, 6280-000104, 6280-000103, 6280-000102, 6280-000101, 6280-000100, 6280-000600, 6280-000276, 6280-001374, 6280-000195, 6280-000192, 6280-000189, 6280-000187, 6280-000178, 6280-000194, 6280-001820, 6280-001805, 6280-001444, 6280-001225, 6280-000850, 6280-000725, 6280-000694, 6280-000553, 6280-000549, 6280-000548, 6280-000330, 6280-000317, 6280-000188, 6280-000182, 6280-000290, 6280-001818, 6280-001765, 6280-001602, 6280-001601, 6280-001450, 6280-001449, 6280-001377, 6280-001376, 6280-001178, 6280-001049, 6280-000796, 6280-000743, 6280-000491, 6280-000356, 6280-000295, 6280-000293, 6280-000292, 6280-000291, 6280-000231, 6280-000196, 6280-000061, 6280-000060, 6280-000473, 6280-000232, 6280-000062, 6280-001319, 6280-001313, 6280-000384, 6280-000199, 6280-000198, 6280-000065, 6280-000064, 6280-000237, 6280-000406, 6280-000401, 6280-000362, 6280-000270, 6280-000400, 6280-001245, 6280-001101, 6280-000873, 6280-000871, 6280-000870, 6280-000861, 6280-000860, 6280-000815, 6280-000554, 6280-000527, 6280-000499, 6280-000416, 6280-000371, 6280-000197, 6280-000523, 6280-000063

Distribution

Worldwide distribution - US Nationwide distribution in the states of NY, AR, CO, FL, NJ, AL, KY, TN, SC, WA, OK, MN, PA, GA, VA, CA, OH, IA, KS, MI, IN, TX, MT, MO, ME, LA, IL, SD, MA, OR, ID, MD, AZ, AK and the countries of MB, ON, FR, QC, BC, IS, CT, RM.

Quantity

529