TS ROC Crown Component 4UR (5/pk), TS ROC Crown Component 5UR (5/pk), TS ROC Crown Component 6UR (5/pk), TS ROC Crown Component 7UR (5/pk), TS ROC Crown Component 8UR (5/pk), TS ROC Crown Component 4LR (5/pk), TS ROC Crown Component 5LR (5/pk), TS ROC Crown Component 6LR (5/pk), TS ROC Crown Component 7LR (5/pk), TS ROC Crown Component 8LR (5/pk), TS ROC Crown Component 4UL (5/pk), TS ROC Crown Component 5UL (5/pk), TS ROC Crown Component 6UL (5/pk), TS ROC Crown Component 7UL (5/pk), TS ROC Crown Component 8UL (5/pk), TS ROC Crown Component 4LL (5/pk), TS ROC Crown Component 5LL (5/pk), TS ROC Crown Component 6LL (5/pk), TS ROC Crown Component 7LL (5/pk), TS ROC Crown Component 8LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
Enforcement
- Recall Number
- Z-1511-2015
- Event ID
- 70787
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Specialty Appliance Works, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 29, 2015
- Initiation Date
- March 9, 2015
- Classification Date
- April 22, 2015
- Termination Date
- January 12, 2017
- Address
- 4905 Hammond Industrial Dr, Cumming, GA, 30041-3914, United States
Description
TS ROC Crown Component 4UR (5/pk), TS ROC Crown Component 5UR (5/pk), TS ROC Crown Component 6UR (5/pk), TS ROC Crown Component 7UR (5/pk), TS ROC Crown Component 8UR (5/pk), TS ROC Crown Component 4LR (5/pk), TS ROC Crown Component 5LR (5/pk), TS ROC Crown Component 6LR (5/pk), TS ROC Crown Component 7LR (5/pk), TS ROC Crown Component 8LR (5/pk), TS ROC Crown Component 4UL (5/pk), TS ROC Crown Component 5UL (5/pk), TS ROC Crown Component 6UL (5/pk), TS ROC Crown Component 7UL (5/pk), TS ROC Crown Component 8UL (5/pk), TS ROC Crown Component 4LL (5/pk), TS ROC Crown Component 5LL (5/pk), TS ROC Crown Component 6LL (5/pk), TS ROC Crown Component 7LL (5/pk), TS ROC Crown Component 8LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.
Product No. 91720, 91721, 91722, 91723, 91724, 91725, 91726, 91727, 91728, 91729, 91730, 91731, 91732, 91733, 91734, 91735, 91736, 91737, 91738, and 91739
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
14,428 Universal Nuts