Nuclear gamma cameras (ADAC Vertex Plus, ADAC Solus, ADAC Vertex V60 & ADAC Vertex Classic) Philips Medical System, Cleveland, OH 44143. Used to produce images depicting the anatomical distribution of single photon and positron emitting radioisotopes with the human body for interpretation by medical personnel.

Recall: Z-1510-2014 · Reported May 7, 2014

Enforcement

Recall Number
Z-1510-2014
Event ID
67901
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 7, 2014
Initiation Date
March 10, 2014
Classification Date
April 25, 2014
Termination Date
August 4, 2015
Address
595 Miner Rd, Cleveland, OH, 44143-2131, United States

Description

Nuclear gamma cameras (ADAC Vertex Plus, ADAC Solus, ADAC Vertex V60 & ADAC Vertex Classic) Philips Medical System, Cleveland, OH 44143. Used to produce images depicting the anatomical distribution of single photon and positron emitting radioisotopes with the human body for interpretation by medical personnel.

Reason

Slippage of the radius drive belt in the relative 180 degree orientation of the superior positioned detector head will allow the detector head to drift down and potentially come in direct contact with the patient.

Code Info

ADAC Vertex Plus & ADAC Solus System Code: 88297, Serial #'s: 11, 1006, 1008, 1027, 1092, 1097, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 6052, 6261, 8002, 10252, 10630, 11090, 85247, 513519, 513520, 940384, 9205345, 9408136, 9503084, 9506016, 9511123, 9512009, 9512018, 9601012, 9601015, 9601016, 9601024, 9601036, 9602161, 9602254, 9603336, 9604016, 9604021, 9605024, 9605026, 9605037, 9606211, 9607015, 9607024, 9608031, 9608046, 9608049, 9608052, 9609063, 9609081, 9610058, 9611035, 9611077, 9611082, 9612072, 9702095, 9702150, 9703154, 9703165, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708076, 9708078, 9708084, 9708088, 9708227, 9709022, 9709061, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711271, 9711291, 9801047, 9801227, 9801232, 9801297, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803303, 9803312, 9803316, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9805376, 9806014, 9806363, 9806377, 9806388, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901162, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905003, 9905012, 9905659, 9905667, 9906670, 9907016, 9907019, 9909011, 9909161, 9910015, 9910017, 9910031, 9911458, 9911698, 9911702, 9911711, 9912181, 9912463, 98066002, 802080322, 0002032, 0007001M, 0008009B, 0011002B, 0101002C, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 130234-01NVERTEX, 5934V, 6090VM, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 9054VM/V, 9056VM/V, 9130VM, 9806373A, 9810428ADAC#7973VM, 9894K, AD1261, AMT00848VM, B0006129, B0007273, B0008001, B0008003, B0008012, B0008021, B0008030, B001200212, B0102114, B0104125, B0104128, B0105138, B0106169, B01070191, B01070192, B0107179, B01080199, B01080204, B01080207, B01090227, B01090228, B01100231, B01100232, B01100233, B01100236, B01100237, B01100238, B01100241, B01100242, B01110251, B01110252, B02050071, B02050262, B02050263, B02050265, B02060167, B02060270, B02060279, B0206027S, B02070289, B02070290, B02070291, B02070297, B02070300, B0207296, B02080318, B0208312, B02090327, B02090334, B02100345, B02110348, B02120352, B03010355, B03010359, B03020048, B03080380, C0006050, C0006051, C0007060, C0008070, C0009075, C0101099, C0102103, C0102108, C0104118, C0107145, C01080152, C01080153, C01080154, C02100218, C02110220, C02120223, C03010227, C03050230, C03060193, CNF, M@10458, M@10482, M@10547, M@105561, M@10573, M@10582, M@10589, M@10591, M@10676, M@10772, M@10837, M@10839, M@10848, M@10850, M@10871, M@10878, M@10882, M@10927, M@10937, M@10967, M@10979, M@11015, M@11020, M@11036, M@11103, M@11109, M@11140, M@11178, M@11184, M@11381, M@11398, M@11423, M@11498, M@11525, M@11571, M@11590, M@11609, M@11644, M@11881, M@11939, M@11949, M@11957, M@11972, M@12060, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296, M@13466, M@13498, M@13543, M@13549, M@13663, M@13665, M@13688, M@13691, M@13748, M@13813, M@14062, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@6002J, M@6012J, M@6075VM, M@6076K, M@6077K, M@6108VA, M@6123VA, M@6143VA, M@6160VM, M@6183VA, M@6244VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6344VM, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6366K, M@6368J, M@6370VA, M@6371VM, M@6375J, M@6376J, M@6460VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6769K, M@6771VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6912K, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6946VM, M@6947J, M@6985J, M@7003VA, M@7004VA, M@7005VA, M@7017VM, M@7024VM, M@7034K, M@7042VM, M@7053VM, M@7070VM, M@7073VM, M@7085K, M@7086J, M@7095VM, M@7096VM, M@7099K, M@7100VM, M@7101VM, M@7102VM, M@7108VM, M@7109VM, M@7151VM, M@7166K, M@7171VM, M@7172K, M@7181J, M@7190VM, M@7198VA, M@7205K, M@7215K, M@7216VM, M@7217VM, M@7221VA, M@7222VA, M@7229VM, M@7233VM, M@7237VM, M@7245J, M@7246VM, M@7247VM, M@7250VM, M@7253K, M@7300J, M@7308VM, M@7358VA, M@7376VM, M@7380VM, M@7384VM, M@7390VM, M@7393VM, M@7394VM, M@7410VM, M@7420VM, M@7463VM, M@7464VA, M@7470VM, M@7490VM, M@7494K, M@7501VM, M@7503VM-FU1, M@7504VM, M@7505VM1, M@7514VM, M@7515VM, M@7536VM, M@7537VM, M@7538VM, M@7556VM, M@7557VM, M@7569VM, M@7572VM, M@7591VM, M@7592VM, M@7603K, M@7605VM, M@7606VM, M@7623K, M@7624K, M@7626K, M@7627K, M@7628VA, M@7638VM, M@7643VM, M@7654VM, M@7664VM, M@7665K, M@7673VM, M@7674K, M@7679VM, M@7682K, M@7685VM, M@7691VM, M@7695VM, M@7697VM, M@7723VM, M@7729VM, M@7733VM, M@7739K, M@7740VM, M@7760VM, M@7761VM, M@7770VM, M@7776VM, M@7789K, M@7801VM, M@7811VM, M@7812VM, M@7819VA, M@7823K, M@7824VM, M@7830K, M@7855J, M@7857VM, M@7860VM, M@7861VM, M@7874VM, M@7877VM, M@7918K, M@7924K, M@7925K, M@7929VM, M@7930VM, M@7932VM, M@7946J, M@7957VM, M@7961VM, M@7962VM, M@7963K, M@7965VM, M@7968VM, M@7975K, M@7987VM, M@7995VM, M@7997VM, M@8004VM, M@8018J, M@8021VM, M@8044VM, M@8064VM, M@8076K, M@8089VM, M@8096VM, M@8105VM, M@8124VA, M@8146VM, M@8155VM, M@8178VM, M@8203VM, M@8219VA, M@8240K, M@8241VM, M@8247K, M@8248VM, M@8252VM, M@8284K, M@8301K, M@8302VM, M@8306K, M@8337VM, M@8356K, M@8380VM, M@8394K, M@8399VM, M@8436K, M@8462VM, M@8484VM, M@8491K, M@8510VM, M@8528VM, M@8537VM, M@8540K, M@8544K, M@8551VM, M@8561K, M@8606K, M@8609VM, M@8651VM, M@8683VM, M@8686K, M@8692VM, M@8693VM, M@8726VM, M@8738K, M@8759K, M@8771VM, M@8781VM, M@8797K, M@8799RV, M@8800VM, M@8803K, M@8812VM, M@8814VM, M@8817RK, M@8876K, M@8882K, M@8884K, M@8888K, M@8897K, M@8903K, M@8906RK, M@8920VM, M@8922VM, M@8935K, M@8936K, M@8942K, M@8943K, M@8948RV, M@8955K, M@8966K, M@8967K, M@8990K, M@8996VM, M@900K, M@9032K, M@9041K, M@9046K, M@9049VM, M@9063K, M@9080K, M@9081K, M@9085K, M@9091VM, M@9107RV, M@9118K, M@9129K, M@9131K, M@9140K, M@9166VM, M@9185K, M@9186K, M@9191K, M@9194VM, M@9221VM, M@9222VM, M@9231K, M@9247RV, M@9252VM, M@9267VM, M@9283K, M@9293VM, M@9308K, M@9309RV, M@9316VM, M@9327K, M@9337K, M@9338VM, M@9340VM, M@9344K, M@9349K, M@9360VM, M@9418VM, M@9426K, M@9434K, M@9442K, M@9452K, M@9471RV, M@9476K, M@9479K, M@9479K, M@9480K, M@9502K, M@9515VM, M@9522K, M@9525K, M@9532VM, M@9542VM, M@9543K, M@9562K, M@9563K, M@9566K, M@9583VM, M@9590K, M@9594K, M@9598K, M@9607VM, M@9609K, M@9622K, M@9639K, M@9646K, M@9653VM, M@9672VM, M@9712K, M@9724VM, M@9730K, M@9765K, M@9769K, M@9799K, M@9816K, M@9831VM, M@9834K, M@9837K, M@AMT1258, M@MCD6416VM, M@MCD7036VM, M@O1317, M@R372820, M@S5195VA, M@S5365-V, M@S5820VM, M@S5907V11, M@S6155VA, M@S7115VM, M@S7156VM, M@S7223K, M@S7559VA, M@S7571VM, M@S7817K, M@S7856K, M@S7864K, M@S7904VM, M@S7941VA, M@S8687K, N@9369VM, R03010363, S0011003, S6235VM, UNK-6597VM, UNK-7590VM, UNK-8097J1, UNK-8533VM1, UNK-8860VM, UNK-9868K, V0007070, V0012099, V0104109, V0104110, V0105112, V0106121, V0106123, V0106124, V01070133, V01090146, V01100151, V01100156, V02050166, V02070174, V02070179, V02080190, V02080194, V02080196, V02080200, V03060224, & V03070229. ADAC Vertex V60 System Code: 882030; Serial #'s: 10588, 9601004, V0005062, V03080235, V03080236, & V2070177. ADAC Vertex Classic System Code: 890122; Serial #'s: 64, 103, 500, 3334, 5005, 9308007, 9309020, 9310032, 9401059, 9401061, 9403076, 9404106, 9407122, 9410002, 9411006, 9411016, 9501210, 9502067, 9502070, 9503003, 9503079, 9503109, 9504019, 9504099, 9505021, 9505124, 9506140, 9506178, 9508171, 9508184, 9508193, 9509206, 9510222, 9510226, 9601146, 9602249, 9608042, 9609084, 9901004, 9901135, 0407V, 0682V, 0795V, 0822VP2, 20, 5405V, 93-10660-1292, 9408128, 9411007, 9502086, 9508169, 9509198, 9509199, B0106154, B02120033, M@10229AMT, M@207JD, M@4324V, M@4404V, M@4431V, M@4432V, M@4480V, M@4623V, M@4630V, M@4639V, M@4686V, M@4692V, M@4835V, M@4858V, M@4893V, M@4959V, M@5032V, M@5033V, M@5038V, M@5105V, M@5145V, M@5158V, M@5189V, M@5280V, M@5289V, M@5338V, M@5360V, M@5398V, M@5407V, M@5412V, M@5431V, M@5439V, M@5467V, M@5474V, M@5476V, M@5497V, M@5498V, M@5515V, M@5526V, M@5590V, M@5617V, M@5627V, M@5661V, M@5662V, M@5663V, M@5664V, M@5665V, M@5681V, M@5688V, M@5689V, M@5751V, M@5752V, M@5782V, M@5811V, M@5835V, M@5869V, M@5900V, M@5913V, M@5915V, M@5960V, M@5970V, M@6004V, M@6007V, M@6010V, M@6056V, M@6900V, M@9304RV, M@9374RV, M@9645RV, M@MVS5236V, M@MVS5852V, M@O1325 & UNK-AMT1238A.

Distribution

Worldwide Distribution -- USA, and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Denmark, Egypt, France, Germany, Hong Kong, Ireland, Italy, Korea, Lebanon, Malaysia, Netherlands, Norway, Pakistan, Portugal, Russia, Singapore, Spain, Sweden, Thailand, & United Kingdom.

Quantity

1,043 cameras