FDA Enforcement Class II Terminated

Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was distributed with the following sheath/probe: 0250070451-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Spatula-Tip, 0250070452-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, J-TIP, 0250070453-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, L-TIP, 0250070455-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Needle Tip

Recall: Z-1509-2020 · Reported March 25, 2020

Enforcement

Recall Number
Z-1509-2020
Event ID
85100
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 25, 2020
Initiation Date
September 12, 2019
Classification Date
March 17, 2020
Termination Date
October 31, 2022
Address
5900 Optical Ct, San Jose, CA, 95138-1400, United States

Description

Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was distributed with the following sheath/probe: 0250070451-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Spatula-Tip, 0250070452-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, J-TIP, 0250070453-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, L-TIP, 0250070455-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Needle Tip

Reason

Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.

Code Info

Lot Numbers: 010818-02, 010818-05, 010818-06, 010818-07, 010818-08, 010818-09, 040816-01, 040816-02

Distribution

OUS: Netherlands, Spain, Switzerland, Colombia, Canada, Japan, Taiwan,, Puerto Rico, South Africa, and South Korea