FDA Enforcement
Class II
Ongoing
Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007
Recall: Z-1508-2025
·
Reported April 9, 2025
Enforcement
- Recall Number
- Z-1508-2025
- Event ID
- 96414
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- PYRAMES INC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Visit
- Report Date
- April 9, 2025
- Initiation Date
- February 24, 2025
- Classification Date
- April 2, 2025
- Address
- 21730 Stevns Crk Blvd Ste 201a, Cupertino, CA, 95014-1171, United States
Description
Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007
Reason
Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.
Code Info
Software version: 2.2.3 Part/UFI-DI/Lot: 54-0002/10860007282925/24020902, 24040302, 23103002; 51-0007/10860007282901/24011702, 24111301
Distribution
US distribution to states of: MA and CA
Quantity
35