FDA Enforcement Class II Ongoing

HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25

Recall: Z-1506-2022 · Reported August 17, 2022

Enforcement

Recall Number
Z-1506-2022
Event ID
90224
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Guangdong Haiou Medical Apparatus Co., Ltd
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 17, 2022
Initiation Date
June 10, 2022
Classification Date
August 5, 2022
Address
Nanyuan Industrial, Zone, North Liusha, Puning, N/A, China

Description

HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25

Reason

Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user

Code Info

Lot Number: 20IB0 Exp. Date: 2025/09/04

Distribution

US Nationwide distribution in the state of CA.

Quantity

4,050,000 units