FDA Enforcement Class II Terminated

Giraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

Recall: Z-1506-2017 · Reported March 29, 2017

Enforcement

Recall Number
Z-1506-2017
Event ID
76283
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ohmeda Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 29, 2017
Initiation Date
November 18, 2016
Classification Date
March 20, 2017
Termination Date
July 13, 2018
Address
8880 Gorman Rd, Laurel, MD, 20723-5800, United States

Description

Giraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

Reason

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Code Info

See Attached Spreadsheat

Distribution

Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,

Quantity

5,993 (5,080 US; 913 OUS)